Instruction Manual Alpha-COND 550
23
measuring range is selected to ‘0.0 to 200.0 mS/cm’.
Press or key to set the line resistance of the sensor cable
connected to the monitor.
Press ENT key to confirm your selection.
4
Reset to factory defaults: The display shows ‘nO dEF’.
Press or key to select:
nO dEF : Not default reset of the monitor. User configured values
remain active, when confirmed by pressing ENT key.
FCt dEF: Reset all settings of the monitor to factory defaults, when
confirmed by pressing ENT key.
CAL dEF: Reset all calibration settings of the monitor to factory
defaults, when confirmed by pressing ENT key.
NOTE: Refer section 6 for factory/calibration default settings
Press ENT key to confirm your selection.
- If ‘nO dEF’ was selected, the monitor reverts to Cnfg screen.
Press or key to access other setup screens or press
and key simultaneously (escape) to return to measurement
mode.
- If FCt dEF’ or ‘CAL dEF’ was selected, the monitor performs the
selected reset and returns to measurement mode.
Press or key to access other setup screens or press and key
simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press and keys simultaneously
(escape). The monitor returns to the first screen configuration settings (Cnfg).