5.2 Temperature Coefficient Settings
Setting the temperature coefficient to match your process/calibration liquids allows
monitor to achieve conductivity readings with higher accuracy.
ENT
SETUP
HOLD
1
SETUP
HOLD
2
ENT
SETUP
HOLD
3
ENT
4
ENT
SETUP
HOLD
2
%
SETUP
HOLD
%
ENT
11
From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press or key to select temperature coefficient settings
screen (tC).
Press ENT key to access temperature coefficient settings (tC).
2
Select type of temperature coefficient: The upper display shows the
last configured temperature coefficient type (Lin or PUr). Press or
key to select the desired type based on type of your process liquid.
PUr :
Pure water temperature coefficient: Select this if your process
liquid is pure water or ultra pure water.
Lin :
Linear temperature coefficient: Select this for all other
process liquids
Press ENT key to confirm your selection.
If PUr was selected:
The monitor reverts back to tC screen
If Lin was selected:
3
Select temperature coefficient for process liquid: The lower
display shows ‘P.tC’. The upper display shows last configured
temperature coefficient for process liquid (if any). The default value
is 2.1%. Press or key to select the desired value. Allowable
range: 0 to 10%.
Press ENT key to confirm your value.
4
Select temperature coefficient for calibration liquid: The lower
display shows ‘C.tC’. The upper display shows last configured
temperature coefficient for calibration liquid (if any). The default
value is 2.1%. Press or key to select the desired value.
Allowable range: 0 to 10%. Press ENT key to confirm your value.
The monitor reverts to tC screen.
Press or key to access other setup screens or press and key
simultaneously (escape) to return to measurement mode.