Alpha COND 550 Conductivity Monitor mS/µS/ºC/ºF meter
Preface This manual serves to explain the use of the Alpha-COND 550 Industrial Conductivity Monitor. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover many anticipated applications of the Alpha-CON 550 monitor. If you have doubts in the use of the instrument, please do not hesitate to contact the nearest Alpha Authorised Distributor.
TABLE OF CONTENTS 1. Introduction ...........................................................................................................1 1.1 1.2 1.3 1.4 Before You Begin ................................................................................................................... 1 Intended Use .......................................................................................................................... 1 Safety Instructions............................................................
Instruction Manual Alpha-COND 550 1. Introduction 1.1 Before You Begin We thank you for purchasing the Alpha COND 550 monitor. The construction of the COND 550 monitor employs leading edge technology and complies with safety regulations currently in force. Notwithstanding this, improper use could lead to hazards for the user or a third-party, and/or adverse effects on the plant or other equipment.
Instruction Manual Alpha-COND 550 1.3 Safety Instructions - The Alpha COND 550 monitor should be installed and operated only by personnel familiar with the instrument and who are qualified for such work. - A defective CON monitor must neither be installed nor put into service. - The Alpha COND 550 must only be operated under the specified operating conditions (see section 6). - The Alpha COND 550 must not be repaired by the customer. - No modifications to the Alpha COND 550 are allowed.
Instruction Manual Alpha-COND 550 2. Getting Started 2.1 Description of Instrument The Alpha COND 550 monitor is used for measuring conductivity and temperature values. The monitor can be used for applications such as water treatment and controlling, galvanic-decontamination, chemical processing, food processing, clean or wastewater control and neutralization processes.
Instruction Manual Alpha-COND 550 2.
Instruction Manual Alpha-COND 550 2.3 Connecting Peripherals 2.3.1 Connection Terminals Remove Back Cover: Remove the screws from the four corners at the back of the conductivity monitor. Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below. Connectors: • J11 – 24V DC power connector • J8 – 9V DC power connector (1.
Instruction Manual Alpha-COND 550 2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors The conductivity monitor supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J7) as described below. Remove Back Cover: Remove screws from the four corners at the back of the conductivity monitor. Remove the back cover.
Instruction Manual Alpha-COND 550 2.3.3 Connecting Conductivity Probe Any industrial conductivity probe (2-cell or 4-cell) can be connected to the conductivity monitor. 2-Cell Conductivity Probe: 1. Short Pin 1 & 2 of J10 connector using a jumper wire. 2. Connect CON low input wire (white) to Pin 2 of J10 connector. 3. Connect CON high input wire (black) to Pin 3 of J10 connector. 4. Short Pin 3 & 4 of J10 connector using a jumper wire. 4-Cell Conductivity Probe: 1.
Instruction Manual Alpha-COND 550 2.4 Installation 2.4.1 Mechanical Dimensions lpha COND550 ESC CAL ENT Conductivity Monitor 2.4.
Instruction Manual Alpha-COND 550 2.4.3 Panel Mount 1 Prepare panel cut-out of 92.0 mm X 92.
Instruction Manual Alpha-COND 550 2.5 Display & Keypad 2.5.1 Display Overview The Liquid Crystal Display (LCD) of Alpha COND 550 conductivity monitor has two alpha-numerical displays (Upper and a Lower). • Upper display: Measured conductivity value is displayed when the monitor is in normal operation (measurement) mode. • Lower display: Measured temperature value is displayed when the monitor is in normal operation (measurement) mode.
Instruction Manual Alpha-COND 550 Units of Measurement Indicators mS Conductivity in Milli Siemens µS Conductivity in Micro Siemens R. Line Resistance in Ohms (Refer Section 5.5) % Temperature Coefficient (Refer Section 5.2) ºC Temperature in Celsius (Refer Section 5.3) ºF Temperature in Fahrenheit (Refer Section 5.3) 2.5.
Instruction Manual Alpha-COND 550 3. Operation 3.1 Measurement mode When the conductivity monitor is powered on, the display shows all the LCD segments briefly, and then automatically enters into the measurement mode. MEAS µS READY 25.0 °C ATC The mode indicator ‘MEAS’ at the top of the display indicates that the conductivity monitor is in measurement mode. The upper alpha-numerical display shows the measured conductivity value, while the lower display shows the temperature value.
Instruction Manual Alpha-COND 550 3.
Instruction Manual Alpha-COND 550 4. Calibration Mode 4.1 About Calibration Calibration should be carried out each time a new electrode is attached to the monitor or when you suspect that the monitor/electrode is out of calibration. The monitor allows you to perform temperature calibration and conductivity calibration.
Instruction Manual Alpha-COND 550 4.2 Calibration The monitor requires one-point calibration. Conductivity calibration is always carried out in the specific measuring range selected for the monitor. Make sure, the monitor is set to desired measuring range before you enter calibration mode (Refer Section 5.4 for selecting measuring ranges). Select a suitable standard conductivity solution so that its conductivity value is within 10% to 100% of the full scale of the selected measuring range of the monitor.
Instruction Manual Alpha-COND 550 NOTES: • The acceptable calibration window is ±40% of the displayed (default) value. For example, if the display is 1000 µS, the values to which it can be adjusted is 600 to 1400 µS. • If the conductivity reading of the solution is not within 10% to 100% of the full scale of the selected measuring range, calibration error occurs; LCD displays “ERR” and flash electrode annunciator.
Instruction Manual Alpha-COND 550 5. Setup Mode 5.1 Enter Setup mode The setup mode allows you to customize the settings of the conductivity monitor to suite your requirements. While in measurement mode, press the ENT key to access setup mode. LCD shows ‘SETUP’ mode indicator and the first page of setup (rAng – Range settings). Press ▲ or ▼ key to access other pages of the setup mode.
Instruction Manual Alpha-COND 550 5.2 Temperature Coefficient Settings Setting the temperature coefficient to match your process/calibration liquids allows monitor to achieve conductivity readings with higher accuracy. 1 SETUP HOLD 4 3 2 ENT SETUP SETUP ENT SETUP % HOLD HOLD ENT % ENT HOLD 2 SETUP ENT HOLD 1 From measurement mode press ENT key to enter setup mode as described in section 5.1. The LCD shows the first screen of setup mode (rAng).
Instruction Manual Alpha-COND 550 5.3 Temperature Settings Automatic Temperature Compensation (ATC): The conductivity value of a sample is affected by temperature. Use ATC feature of the monitor to compensate for conductivity changes when the temperature of the sample or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) to the monitor if an integrated conductivity/Temperature probe is not available.
Instruction Manual Alpha-COND 550 If ATC enabled (ATC On): 4 Selecting temperature offset: The upper display shows the last configured temperature offset value (if any), otherwise the default is zero. The lower display shows currently measured temperature reading (including last configured offset value). LCD shows ‘ATC’ annunciator in lower-right corner. Place a thermometer, which is known to be accurate, in your sample or process liquid. Make sure your temperature probe is placed in the same liquid.
Instruction Manual Alpha-COND 550 5.4 Measuring Range Settings The conductivity monitor has 5 different measuring ranges. You need to select a range which is the most suitable for conductivity reading of your applications/samples. The factory default is 2000 µS (0 to 2000 µS/cm). The measuring range you select here will affect the way the monitor behaves in the measurement mode and calibration mode.
Instruction Manual Alpha-COND 550 5.5 Configuration Settings Configuration settings let you select between 2-cell & 4-cell and reset the monitor to factory defaults. 1 SETUP HOLD 3 2 SETUP SETUP ENT ENT HOLD HOLD 4 ENT SETUP ENT HOLD This screen appears only when 2-cell is selected and measuring range is set to 0.0 to 200.0 mS/cm 2 SETUP R.
Instruction Manual Alpha-COND 550 measuring range is selected to ‘0.0 to 200.0 mS/cm’. Press ▲ or ▼ key to set the line resistance of the sensor cable connected to the monitor. Press ENT key to confirm your selection. 4 Reset to factory defaults: The display shows ‘nO dEF’. Press ▲ or ▼ key to select: nO dEF : Not default reset of the monitor. User configured values remain active, when confirmed by pressing ENT key.
Instruction Manual Alpha-COND 550 5.6 Viewing Electrode Properties Each time you calibrate the conductivity sensor, the monitor re-calculates the calibration factor of the sensor and shown in the LCD at the end of calibration. The setup mode allows you to view the last saved calibration factor and cell constant at any time. 2 1 SETUP HOLD 1 ENT SETUP ENT HOLD SETUP ENT HOLD From measurement mode press ENT key to enter setup mode as described in section 5.1.
Instruction Manual Alpha-COND 550 6. Technical Specifications General Specification Conductivity Measuring Range: 0.00 - 20.00 µS/cm 0.0 - 200.0 µS/cm 0- 2000 µS/cm 0.00 - 20.00 mS/cm 0.0 - 200.
Instruction Manual Alpha-COND 550 Calibration Default Settings (CAL dEF) Conductivity Calibration factor Reset to 1.000 Factory Default Settings (FCt dEF) Conductivity Calibration factor Cell constant Cell type Measuring range Line resistance Temperature Unit of measurement Temperature compensation Temperature coefficient Temperature coefficient type MTC (ATC Off) Reset to 1.000 1.00 2-cell 0 to 2000 µS/cm 0.000 Ω oC ATC On 2.1% (Both process & calibration temp.
Instruction Manual Alpha-COND 550 7. List of Accessories Monitor Replacement and Accessories Item Description Alpha CON 550 Monitor Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 0.1 Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 1.0 Conductivity 2 Cell type probe, 0.1 - 200μS; Cell constant, K=0.1 with integrated Pt 100, Material SS316 and 25ft cable (open-ended) Conductivity 2 Cell type probe, up to 200 mS; Cell constant, K=1.
Instruction Manual Alpha-COND 550 8.
Instruction Manual Alpha-COND 550 9. General Information 9.1 Warranty Eutech Instruments warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and not the result of abuse or misuse within the warranty period, please return by freight pre-paid and amendment will be made without any charge. Eutech Instrument Customer Service Dept.
Instruction Manual Alpha-COND 550 10. Appendices 10.1 Appendix 2 Conductivity of various aqueous solutions At Temperature 25 oC /77 oF Solution Ultra-pure Water Conductivity 0.055 µS/cm Resistivity 18.18 MΩ-cm 0.05 - 1 µS/cm 1 - 18 MΩ-cm 0.5 µS/cm 2 MΩ-cm 0.1 - 10 µS/cm 0.1 - 10 MΩ-cm 1 - 80 µS/cm 0.01 - 1 MΩ-cm Mountain Water 10 µS/cm 0.1 MΩ-cm Drinking Water 0.5 - 1 mS/cm 1 - 2 kΩ-cm Waste-water 0.9 - 9 mS/cm 0.1 - 1 kΩ-cm 1.5 mS/cm 0.7 kΩ-cm Brackish Water 1 - 80 mS/cm 0.
Instruction Manual Alpha-COND 550 10.2 Appendix 4 - Abbreviations Used in LCD Abbreviation Description AtC CAL C.ºC CdAt CEL CnFg C.tC dEF FCt HOLD K L.Ad Lin MEAS Or P.ºC P.
NOTES
Water Analysis Instruments North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031 Europe P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832 Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836 www.thermo.com/process © 2009 Thermo Fisher Scientific Inc. All rights reserved.