6.9 Measuring range selection (RANG) sub-function
In this sub-function you select the measuring
range.
1. Select the “RANG” sub-function, then press the ENT key.
2. Selecting the measuring range: The display shows the
currently selected measuring range. Press the or key to
select the requested measuring range.
Measuring ranges: No. Measuring range Resolution
1 0.000 to 2.000 µS/cm 0.001 µS/cm
2 0.00 to 20.00 µS/cm 0.01 µS/cm
3 0.0 to 200.0 µS/cm 0.1 µS/cm
4 0 to 2000 µS/cm 1 µS/cm
5 0.00 to 20.00 mS/cm 0.01 mS/cm
6 0.0 to 200.0 mS/cm 0.1 mS/cm
7 0 to 1000 mS/cm 1 mS/cm
Relative accuracy: ±1 % of full scale reading (±2 % >500 mS/cm)
System accuracy: The effective measuring range and the total system accuracy depend on the
sensor used with the transmitter. Please refer to the respective sensor
documentation.
Press the ENT key to confirm your selection.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
and keys simultaneously (escape).