5.4.3 The Lo SEt Menu
1. Enter this menu by pressing
MODE key twice to the “SEt”
display if you are in measurement
mode.
2. Press ENTER key. If the
password has been set earlier, key
in the password using the method
described in section 4.1. Press
ENTER once to confirm.
3. After you set the correct
password, you see the “CAL” and
“COn” display blinks alternately.
Press
▼ key four times, the display
shows “Lo” “SEt” blinking
alternately.
4. Press ENTER key to access
the Lo SET menu and the display
shows the last Lo SET value or
default (5% of Full Scale). Use
▲
and ▼ keys to change the value of
the Lo SET point.
5. Press ENTER to confirm the
value of the Lo SET point.
6. Press MODE key to exit to
measurement mode. You can
press MODE key (as an ESCAPE
key) to revert to the measurement
value if ENTER is not pressed; the
set value is not stored into
memory.
SETUP
MODE
ENTE
R
Lo
SETUP
MODE
ENTE
R
50
SETUP
MODE
ENTE
R
SEt
NOTE : The Lo SEt
relay activates if the
current Conductivity
reading exceeds Lo
SET point - ½ Lo HYS
value. The Lo SET
value cannot be higher
than the Hi SET value.