Specifications

X-Ray Field in the Long Dimension
IR Image Receptor (Cassette Tray)
VIRC Voltage from the Cassette Sensing Element representing the Image
Receptor size in the Cross Dimension
VIRL Voltage from the Cassette Sensing Element representing the Image
Receptor size in the Long Dimension
IR TRUE
Voltage representing the Presence of a Cassette
VCPC Voltage applied to the Collimator Potentiometers in the Cross Position
VCPL Voltage applied to the Collimator Potentiometers in the Long Position
November 2005 ( 8 - 2 ) Linear MC150-C Collimator
TABLE 1.
The following list is intended to help the installer determine mounting information only, and does
not imply compatibility. See Section 1-0 for compatability information.
MANUFACTURER TUBE HOUSING
FOCAL SPOT
TO
PORT MOUNTING
DISTANCE TO
COLLIMATOR
MOUNTING FLANGE
Eureka Emerald Series
Diamond Series
Sapphire Series
2 - 1/16”
2 - 1/16”
2 - 1/16”
3/8” (.95 cm)
3/8” (.95 cm)
1/4” (.635 cm)
Varian/Eimac B100 2 - 17/64” 11/64” (.434 cm)