Specifications

LIGHT FIELD / X-RAY FIELD CONGRUENCE TEST
X1
METAL MARKER
LIGHT FIELD X-RAY FIELD
Y1 Y2
X2
X1 + X2 must be less than 2% of the SID
Y1 + Y2 must be less than 2% of the SID
FIGURE 3.1
November 2005 ( 3-5 ) Linear MC150-C Collimator