Specifications
2.6.1 120 VAC or 220 VAC INPUT
Connect the three wire cable supplied to the VAC source as follows:
Black - Hot
Wire - Neutral
Green - Ground
Progeny offers two different power supplies to match incoming power of either
120 VAC or 220 VAC depending on power source available. The power supply
part numbers are:
120 VAC transformer: P/N 70-20254
220 VAC transformer, P/N 70-20288
Remove the protective sheet metal cover.
Measure AC power source with a RMS type voltmeter and record reading.
Connect the power source to the transformer tap closest to the power source
voltage read. The taps are numbered as follows for:
120 VAC Transformer, (Power Supply 70-20254)
Primary: 0V, 105, 115, 125 VRMS 47-63 Hz
Secondary: Full load 19, 27 VRMS @ 6-25 Amps
220 VAC Transormer, (Power Supply 70-20288)
Primary: 0V, 210, 230, 250 VRMS 47-63 Hz
Secondary: Full load 19, 27 VRMS @ 6-25 Amps
2.6.2 Route the 8-conductor Collimator cable and plug between the Power Chassis
and Collimator Head as desired. Connect plug B4 inserting it into receptacle and
turning clockwise until it is secure.
FIGURE 2.2
SKIN GUARD INSTALLATION
November 2005 ( 2 - 6 ) Linear MC150-C Collimator