Specifications
2.4 INTERCONNECT WIRING
24 VOLT AC SOURCE ONLY
(Refer to Cabling Outline Figure 2.4)
Connect the unterminated end of the supplied three wire cable and plug to a suitable 24
VAC source with a capacity of 8 Amps minimum. Make the connections as per the
following:
BLACK - HOT (fused at 8 AMPS min.)
WHITE - Neutral
GREEN - Ground (Earth)
Connect the plug end of the cable to the receptacle provided in the rear of the collimator
head.
NOTE: It is the responsibility of the installer to insure that the collimator has been
connected to a source which has been fused for no more than 8 Amps.
2.5 POWER CHASSIS MOUNTING
The power chassis is intended to be mounted on the wall or in an equipment cabinet.
There are holes in the bottom of the enclosure to allow mounting. Follow all local wiring
codes and locate the enclosure in an area that will permit:
• Cable Bend Radius
• Convection Cooling
• Access to Fuses F1 and F2
External connections to the system are made to the free ends of the two cables
permanently attached to the Power Chassis. They include:
• 120 VAC or 220 VAC Power Connection (Depends on Power Supply purchased)
• Power Chassis to Collimator (Plug Connection).
Refer to the Power Chassis drawing (Figure 2.5) for mounting dimensions.
2.6 INTERCONNECT WIRING
(Refer to Cabling Outline Figure 2.6)
All connections are made external to the Power Chassis (refer to Figures 2.5 and 2.6)
November 2005 ( 2 - 5 ) Linear MC150-C Collimator