Specifications
2.0 INSTALLATION
2.1 UNPACKING
Carefully unpack the equipment and check for damage incurred during shipment. Any
damage should be referred to the agency that delivered the product.
2.2 EQUIPMENT SUPPLIED
Refer to Figure 2-1 for component identification
• Linear MC150-C Collimator
• Spacers and mounting hardware
• Interconnect cable
• Packet containing Instruction Manual, Assembler’s Report FD -2579, Returned Goods
Authorization/Service Report
• Skin Guard Kit (Optional)
2.3 COLLIMATOR MOUNTING
2.3.1 SKIN GUARD INSTALLATION
NOTE: Refer to Figure 2.2
1. Remove the Front and back covers, then the outer cover by removing the
twelve (12) 6-32 screws which keep them in place.
2. Mount the two (2) Skin Guard Rails to the outer cover using the four (4)
screws provided in the Skin Guard Kit.
3. Reattach the outer cover, then the front and back covers. After
assembly, test the field lamp timer for proper operation.
2.3.2 Determine the collimator mounting surface to focal spot distance from the data
supplied with the X-Ray tube (do not rely on an inscribed mark on the tube
housing).
Note: The collimator will not perform properly unless the focal spot to upper
swivel ring distance is 2-7/16” (2.44 inches, 62 mm) +/- 1/32” (.031 inches, 1
mm). Be sure to include any permanent mounting plates in the focal spot
to port boss distance stated in the tube manufacturer’s data.
Note: The Linear MC150-C is designed to be used with a lead diaphragm or cone
in the plastic port of the X-Ray tube.
If it is found that lead diaphragms or cones require removal or modification,
consult the factory.
November 2005 ( 2 - 3 ) Linear MC150-C Collimator