Specifications

3. Inherent Filtration and Half-Value Layer:
The Eureka Linear™ collimator has a minimum value of 2.0 mm aluminum
equivalence at 100 kV. This value plus any tube inherent filtration plus any
added filtration must meet the minimum requirements of 21 CFR Sub-Chapter J,
Part 1020.30 (m)(1) Table 1 on beam quality (e.g. minimum HVL at 100 kV must
be 2.7 mm Al.
4. Application:
The intended application is for general purpose radiographic equipment,
including tomographic and chest applications. Maximum tube rating must be 150
kV or less.
5. Installation:
Must be made with supplied hardware, including mounting flange, spacers (as
required), four (4) M6 x 16 bolts, and four (4) 1/4 20 screws equally spaced on
a 9 cm diameter bolt center.
1.8 MAINTENANCE
The Collimator system must be properly maintained to assure both compliance with
BRH regulations and useful life.
Preventive maintenance is to be performed once every twelve months. This includes
inspection and lubrication of both the cassette tray(s) and collimator mechanism. The
collimator mounting ring and locking screw (70-10036 and 26-00752) should be
examined to ensure secure mounting of the collimator. ONLY PROGENY P/N
26-00752 COLLAR LOCKING SCREW SHOULD BE USED.
Checkout should also occur if any of the following conditions occur:
Lamp replacement
Premature electronic component failure
When collimator is removed from tube/housing assembly
When collimator and/or cassette tray have been subjected to external damage
Refer to Section 5.0 for collimator CHECK-OUT procedure, and refer to the cassette
tray manual for tray maintenance.
November 2005 ( 1 -8 ) Linear MC150-C Collimator