Specifications
The landing area of the beam will contain a radiographic image receptor located in a
plane perpendicular to the beam at pre-determined distances from the radiation source
(focal spot).
The size and shape of the image receptor will determine the maximum useful cross-
sectional size and shape of the beam in the plane of the image receptor. The source-to-
image receptor distance (SID) determines the actual shutter opening required to
regulate the beam size and shape in the plane of the image receptor.
1.3 LINEAR SERIES COLLIMATOR FEATURES
1.3.1 SERVICEABILITY
The Linear™ series collimator logic provides a third objective not included with
other similar products - serviceability. This new dimension is incorporated in a
manner which allows a single positioning of the collimator above a table top for
the diagnostic troubleshooting of the logic and collimator functions. All
calibrations are then done by observing the light-field projected onto a test
pattern provided with each collimator.
1.4 GENERAL SPECIFICATIONS - Model LINEAR MC150
The Linear MC150-Ccollimation system from EUREKA includes all features required for
diagnostic excellence...
APPLICATIONS.................. For general purpose mobile X-Ray units and special
purpose radiographic units.
MAXIMUM kVp................... 150 kVp.
OUTER DIMENSIONS........ 10.12” x 6.75” x 6.5” (25.70 cm x 17.15 cm x 16.51
cm
NET WEIGHT..................... 17 lbs (10.2 kg) - approximately
PROJECTION LAMP.......... Quick change, pre-aligned quartz halogen lamp. Light
output more than 160 Lux at 100 cm.
LAMP TIMER...................... Push button type, 25 seconds approximately.
POWER SUPPLY............... 24 VAC +/- 10% at 8 Amps
PROJECTED FIELD SIZE.. Square or rectangular pattern continuously variable
from closed to 43cm by 43cm at 86cm SID
(17” x 17” at 34” SID)
CONE
TRACK.....................
Optional
SWIVEL MOUNT................ Standard
BUCKY LIGHT LINE........... Optional
SKIN GUARDS................... Optional, to be attached for Mobile operation
FILTRATION....................... 2.0 mm (Min.) Aluminum equivalent at 100kVp.
INDICATIONS..................... English and Metric cassette sizes at 40”(100 cm) and
72” (180 cm) SIDs.