EUREKA Progeny, Inc. 1407 Barclay Blvd. Buffalo Grove, IL 60089 Tel. (847) 850-3800 Fax (847) 850-3801 LINEAR MC150-C ™ Installation Operation and Maintenance 00-02-1526 Copyright © 1998 Progeny, Inc. Corp. Rev. H ECN: P1095 November 2005 PROGENY, INC.
Manufacturers of Eureka Collimators and Cables INSTALLATION ADVISORY TO: INSTALLERS, SERVICE PERSONNEL, AND USERS OF X-RAY SYSTEMS COLLIMATOR MOUNTING INFORMATION In order to ensure a safe and secure mounting of this collimator to the x-ray tube housing, the following installation guidelines must be followed: 1. Two different length screws are provided in the cloth bag containing the spacers.
SECTION PAGE 1.0 INTRODUCTION ........................................................................... 1-1 1.1 You have Legal Obligations.................................................................... 1-4 1.2 Background............................................................................................ 1-4 1.3 Collimator Features................................................................................. 1-5 1.4 General Specifications......................................
SECTION PAGE Record Sheet.......................................................................................... 5-21 6.0 THEORY OF OPERATION............................................................ 6-1 6.1 Mechanical Operation............................................................................. 6-3 6.2 Electrical Operation................................................................................. 6-3 7.0 8.0 RENEWAL PARTS LIST................................................
SECTION TABLE PAGE 5.0 5-1 Minimum Beam Quality Requirements......................... 5-4 5-2 Aluminum Equivalentof Primary Beam Total Filtration. 5-5 5-3 Highest Design Operating Range................................ 5-7 5-4 Half Value Layers as a Function of Filtration and Tube 5-8 Potential for Diagnostic Units.......................................
SECTION 1.
This product has been tested by Underwriter’s Laboratories in conformance with standards set forth by UL 2601-1, CAN/CSA-C22.2 No. 601.1-M-90, and IEC 601-2-32. It has been found to comply with these standards and, therefore, bears the above “Recognized Component” symbol for UL and UL-C. UL File No.
Figure 1.1 • • • November 2005 1.0 INTRODUCTION MANUAL OPERATION Press the “lamp” pushbutton to activate the light-field. Adjust the shutters (both longitudinal and cross) to a size not larger than the film to be used. Center light-field over cassette or anatomical area to be exposed.
This manual contains information for the assembly, installation, adjustment, testing and maintenance of the LINEAR series of radiographic/fluoroscopic collimators manufactured by Progeny, Inc. 1.1 YOU HAVE LEGAL OBLIGATIONS The manufacturers of beam limiting devices are required to provide instructions for the assembly, installation, adjustment and testing adequate to assure compliance with applicable provisions of DHHS Performance Standards 21 CFR Sub-Chapter J, Part 1020.
The landing area of the beam will contain a radiographic image receptor located in a plane perpendicular to the beam at pre-determined distances from the radiation source (focal spot). The size and shape of the image receptor will determine the maximum useful crosssectional size and shape of the beam in the plane of the image receptor. The source-toimage receptor distance (SID) determines the actual shutter opening required to regulate the beam size and shape in the plane of the image receptor. 1.
1.5 ADVANCED FEATURES The Linear MC150-C Collimator system also incorporates features required for diagnostic convenience... 1.6 … Swivel mount standard for angulated positioning. … All electronic components are standardized types available at major electronic suppliers. Use of standardized, straight-forward logic allows troubleshooting and repair with general electronic experience. … A spare lamp is provided inside the lamp housing and is easily replaced by the owner/operator.
Mechanical-Electrical Warning for Diagnostic X-Ray Systems All of the moveable assemblies and parts of X-Ray equipment should be operated with care. Only properly trained and qualified personnel should be permitted access to any internal parts. Live electrical terminals are deadly; be sure line disconnect switches are opened and other appropriate precautions are taken before opening access doors, removing enclosure panels, or attaching accessories.
3. Inherent Filtration and Half-Value Layer: The Eureka Linear™ collimator has a minimum value of 2.0 mm aluminum equivalence at 100 kV. This value plus any tube inherent filtration plus any added filtration must meet the minimum requirements of 21 CFR Sub-Chapter J, Part 1020.30 (m)(1) Table 1 on beam quality (e.g. minimum HVL at 100 kV must be 2.7 mm Al. 4. Application: The intended application is for general purpose radiographic equipment, including tomographic and chest applications.
SECTION 2.
MC150-C COLLIMATOR INTERCONNECT CABLE MC150-C MANUAL SPACER KIT SKIN GUARD KIT (OPTIONAL) COMPONENT IDENTIFICATION FIGURE 2.
2.0 INSTALLATION 2.1 UNPACKING Carefully unpack the equipment and check for damage incurred during shipment. Any damage should be referred to the agency that delivered the product. 2.2 EQUIPMENT SUPPLIED Refer to Figure 2-1 for component identification • Linear MC150-C Collimator • Spacers and mounting hardware • Interconnect cable • Packet containing Instruction Manual, Assembler’s Report FD -2579, Returned Goods Authorization/Service Report • Skin Guard Kit (Optional) 2.3 COLLIMATOR MOUNTING 2.3.
Determine the total thickness of the supplied spacer(s) that must be added to the collimator mounting surface to obtain a focal spot to collimator mounting flange distance of 2-7/16” (2.44 inches, 62 mm) +/- 1/32” (.031 inches, 1 mm). Refer to Figure 2-3. Remove the upper swivel ring from the collimator by removing the 6-32 socket head cap screw and opening the clamp ring.
2.4 INTERCONNECT WIRING 24 VOLT AC SOURCE ONLY (Refer to Cabling Outline Figure 2.4) Connect the unterminated end of the supplied three wire cable and plug to a suitable 24 VAC source with a capacity of 8 Amps minimum. Make the connections as per the following: BLACK - HOT (fused at 8 AMPS min.) WHITE - Neutral GREEN - Ground (Earth) Connect the plug end of the cable to the receptacle provided in the rear of the collimator head.
2.6.1 120 VAC or 220 VAC INPUT Connect the three wire cable supplied to the VAC source as follows: Black Wire Green - Hot Neutral Ground Progeny offers two different power supplies to match incoming power of either 120 VAC or 220 VAC depending on power source available. The power supply part numbers are: 120 VAC transformer: P/N 70-20254 220 VAC transformer, P/N 70-20288 Remove the protective sheet metal cover. Measure AC power source with a RMS type voltmeter and record reading.
DIMENSIONS FIGURE 2.
FIGURE 2.4 INTERCONNECTION WIRING WHEN USING A 24 VOLT AC SOURCE FIGURE 2.
3 COND. CABLE 20 FT. HIGH ENERGY (NEC) EUREKA SUPPLIED COLLIMATOR 3 COND. 15 FT. 117 VAC INPUT POWER SUPPLY JUNCTION BOX CABLE CONFIGURATION FOR 70-63000 LINEAR MC 150 WHEN USING EUREKA POWER SUPPLY 70-20254 FIGURE 2.
(NO TEXT) November 2005 ( 2 - 10 ) Linear MC150-C Collimator
SECTION 3.
(NO TEXT) November 2005 ( 3-2 ) Linear MC150-C Collimator
3.0 OPERATIONAL CHECK-OUT PROCEDURE 3.1 Operational Check - After the Collimator, Power Chassis and cabling have been installed, apply 24 VAC power and observe the collimator. 3.2 3.1.1 Push the LAMP button and check that the light field lamp remains on for approximately 25 seconds. 3.1.2 While the field lamp is on, be sure the light field can be collimated to required size with the control knobs.
3.2.1 Remove the LINEAR collimator table-top TEST PATTERN #1 from this manual and position it on the table-top with the edges parallel to the table-top edges. Flatten the creases and tap it into position at the corners in a manner that will not damage it upon removal. 3.2.2 Angulate the collimator to 0° horizontal. Position the collimator at a focal spot to TEST PATTERN distance of 100 cm + 0.
LIGHT FIELD / X-RAY FIELD CONGRUENCE TEST X1 METAL MARKER LIGHT FIELD X-RAY FIELD Y1 Y2 X2 X1 + X2 must be less than 2% of the SID Y1 + Y2 must be less than 2% of the SID FIGURE 3.
(NO TEXT) November 2005 ( 3-6 ) Linear MC150-C Collimator
SECTION 4.
CROSS-HAIR WINDOW ADJUSTMENT FIGURE 4.
4.0 ADJUSTMENT AND ALIGNMENT PROCEDURES 4.1 FIELD PROJECTION LAMP AND MIRROR ADJUSTMENT These tests must be performed when the field projection lamp is altered from it’s original position or replaced. These tests must also be performed if the original mirror angle has been altered and if any edge of the developed X-Ray image is outside of the 30.5 cm x 30.5 cm @ 100 cm tolerance marks as defined in Step 3.2. 4.1.1 Steps 3.1 through 3.
4.2 CROSS HAIR WINDOW ADJUSTMENT These procedures are to be performed if the cross hair shadows are not centered to the light field (Reference Figure 4.1). 4.3 4.2.1 Loosen the screws securing the plastic window. 4.2.2 Move the plastic window to align and center the cross hair pattern to the light field (center lines on the test pattern). 4.2.3 Tighten the screws and reassemble the collimator covers. BUCKY CENTERING LIGHT-LINE ADJUSTMENT 4.3.
4.4 LONG AND CROSS SIZE INDICATOR ADJUSTMENT 4.4.1 Using a long allen wrench, loosen the #4-40 set screws securing the indicators to the axles. 4.4.2 Same as above. 4.4.3 Set the indicators to 17” x 17” @ 40” SID then tighten the indicator set screws.
COVER REMOVED Mirror Retractor Lamp/Lightfield Adjustment Longitudinal Adjustment CROSS ALIGNMENT LONGITUDINAL ALIGNMENT 1. Adjust two #6-32 screws for light-field alignment 2. (Option) Add Loctite to set in postion 1. Slightly loosen the two #6-32 screws 2. Position lamp bracket laterally for light-field alignment, left to right 3. Tighten the two #6-32 screws FIGURE 4.2 - LIGHT-FIELD ADJUSTMENT WARNING! THE LAMP AND HEAT DEFLECTORS MAY BE HOT ENOUGH TO CAUSE SEVERE BURNS.
SECTION 5.
FIGURE 5.1 BRN/FDA COMPLIANCE TEST STAND FIGURE 5.
5.0 COMPLIANCE VERIFICATION It is necessary for the assembler to verify compliance. A series of tests, when performed at the time of installation, will indicate compliance with 21 CFR, Sub-Chapter J, Part 1020, Performance Standards. The following tests are from NEMA Standards Publication, No. XR-8-1979 (Test Methods for Diagnostic X-Ray Machines for Use During Initial Installation). For each compliance item, there may be a variety of test methods described.
Those authorized to operate, test, participate in or supervise the operation of the equipment must be thoroughly familiar and comply completely with the currently established safe exposure factors and procedures described in publications such as Sub-Chapter J of Title 21 of the Code of Federal Regulations, “Diagnostic X-Ray Systems and their Major Components,” and the National Council on Radiation Protection (NCRP) No.
Table 5-2 ALUMINUM EQUIVALENT OF PRIMARY BEAM TOTAL FILTRATION Operating Voltage (kVp) Below 50................................... 50-70........................................ Above 70................................... Total Filtration (mm Al Equivalent) 0.5 1.5 2.5 C. PROCEDURE Visually inspect the system and determine the aluminum equivalence of the total filtration in the primary beam.
D. VERIFICATION OF COMPLIANCE Verify that the radiation output in step 2 is greater than or equal to 50 percent of the radiation output in step 1. .03 METHOD III - BRH/FDA COMPLIANCE TEST A. GENERAL The HVL determinations obtained from the following procedures are to be compared with those illustrated in Table 5-1. The HVL in millimeters of aluminum of the system being tested must be greater than or equal to the values shown in Table 5-1. B. EQUIPMENT 1. 2. 3. C.
Below 50 kVp 0.5 1.0 1.5 2.0 Total Added Filtration, mm Al 50 - 70 kVp 1.0 1.5 2.5 3.5 Above 70 kVp 1.5 2.5 3.5 4.5 The recorded data is plotted on semi-log graph paper (Examples A and B, Fig. 5-6) and the half-value is read directly from the graph. D. VERIFICATION OF COMPLIANCE Verify that the half-value layer of the useful beam for a given X-Ray tube potential is not less than the values shown in table 5-1.
TABLE 5-4 Total Filtration mm Al HALF VALUE LAYERS AS A FUNCTION OF FILTRATION AND TUBE POTENTIAL FOR DIAGNOSTIC UNITS* Peak Potential (kVp) 30 40 50 60 70 80 90 100 110 120 1.08 1.70 2.25 2.70 3.06†‡ 3.38 3.68 1.16 1.82 2.42 2.90 3.30†‡ 3.65 3.95 Typical Half-Value Layers (mm Al) 0.5 1.0 1.5 2.0 2.5 3.0 3.5 0.36† 0.55 0.78 0.92 1.02 ... ... 0.47† 0.78 1.04 1.22 1.38 1.49 1.58 0.58 0.95 1.25† 1.49 1.69 1.87 2.00 0.67 1.08 1.42† 1.70 1.95 2.16 2.34 0.76 1.21 1.59† 1.90 2.16 2.40 2.60 0.
XR 8-2.14 VISUAL DEFINITION (RADIOGRAPHIC) OF X-RAY LIGHT FIELD REQUIREMENT - Means shall be provided for visually defining the perimeter of the XRay field.
C. VERIFICATION OF COMPLIANCE For determination of misalignment, compare the edges of the X-Ray field to the edges of the light-field as defined by the outer edges of the metal strips. On each side of the rectangular fields, measure the separation between the X-Ray field and the outside edge on the image of the respective metal strip. Sum these measured separations for opposite sides of the X-Ray field to yield a total misalignment in the length and width dimensions.
FOCAL SPOT BRASS STRIPS X FOCAL SPOT ASSY. IN SLOT TEST STAND 2.5” 13.95” SLID ASSY IN SLOT 6 TABLETOP S FIGURE 5-4 CALCULATION EXAMPLE DOTTED LINE COINCIDES WITH THE OUTER EDGE OF THE MEATAL STRIPS AND IS THE PERIMETER OF THE LIGHT FIELD.
B. EQUIPMENT 1. 2. Plastic cassette with direct printer paper or film. Radio-opaque markers* * Each marker is approximately .080 cm galvanized sheet metal having the dimensions of 4 by 4 cm. C. PROCEDURE 1. Adjust the source assembly and the beam-limiting device so that they are approximately centered over the table and perpendicular to the table top. Then position the beam-limiting device to the SID previously determined and record the indicated value. 2.
E. CALCULATIONS 2.5 S = X X + 13.95 2.5X + (2.5) 13.95 = XS (2.5) 13.95 = XS - 2.5X 34.875 = X (S - 2.5) X = 34.875 S - 2.5 The misalignments are calculated as follows: Length misalignment = L1 + L2 ≤ 2% SID Width misalignment = W 1 + W2 ≤ 2% SID Calculate 2% of the measured SID. Each of the misalignments, length or width, must be less than or equal to 2% of the measured SID for compliance. NEMA Standards 5-15-79 .03 METHOD III - ALTERNATE TEST STAND METHOD A. B. November 2005 GENERAL 1.
C. D. PROCEDURE 1. Align the tube unit and image receptor and set the SID with the normal operating aids (detents, scales, lights, etc.) 2. Load cassette and insert into image receptor. 3. Close shutters to a size smaller than that of the cassette placed into the image receptor. 4. Position the test stand in accordance with the manufacturer’s instructions. 5.
2. 3. B. EQUIPMENT 1. 2. 3. C. The graduated template is utilized to minimize the amount of error introduced into the measurement of the X-Ray field size. The actual versus indicated source-to-image distance (SID) must be determined prior to performing this test. Manufacturer’s recommended test stand. Cassettes and film. Graduated template. PROCEDURE 1. Align the tube unit and image receptor and set the SID with the normal operating aids (detents, scales, lights, etc.) 2.
Digital voltmeter C. PROCEDURE 1. Remove trim covers to gain access to the lamp socket. 2. Verify that the specified lamp is in the socket. 3. With the light-field energized, measure the voltage across the lamp socket terminals. 4. Record the voltage measured. 5. Record the model number, serial number and calibration date of the digital voltmeter. D. VERIFICATION OF COMPLIANCE The voltage recorded shall be within the tolerances specified by the manufacturer.
3. Set the SID to the value determined in the actual versus indicated SID test. 4. Reduce the X-Ray field to approximately 15 by 20 cm. 5. Make an exposure and develop the film. 6. To determine as accurately as possible the corners of the image recorded on the film, locate two points on each of the four sides of the image. Through the two points on each side draw a straight line. These four lines, when extended, intersect making a rectangle which is a close approximation of the actual X-Ray field.
D. 2. Center the film cassette in the cassette tray and insert into position. 3. Adjust the field size to 15 by 15 centimeters or 8 by 8 inches by means of the numerical indicators on the beam-limiting device. 4. Make an exposure and develop film. 5. Measure and record the length and width dimensions of the image. VERIFICATION OF COMPLIANCE The deviation of any of the recorded dimensions must not exceed 2 percent of the SID in Step 1. NEMA Standards 5-15-79 XR 8-2.
4. Center the source assembly over the test stand using the means provided, e.g. bucky light. 5. Bring the source assembly down into firm contact with the spacer. 6. Center the cassette tray with the source assembly using the means provided, e.g. bucky light. 7. Insert the plastic cassette into the slide assembly. Then insert the slide assembly into slot 5 (see Figure 5-2). 8. Center the film cassette in the cassette tray and insert into position.
Prior to performing this test, the magnification factor must be determined in accordance with the X-ray/light field alignment test - Method III. B. EQUIPMENT 1. 2. C. D. Manufacturer’s recommended test stand Cassette with film PROCEDURE 1. Align the tube unit and image receptor and set SID to the value determined in the actual versus indicated SID test. 2. Insert empty 20 by 25 cm cassette into bucky tray. 3. Position test stand in accordance with manufacturer’s instructions. 4.
1. Determination of Half-Value Layer XR8/2.09 2. Visual definition of X-Ray light field XR8/2.14 3. Intensity of light-field XR8/2.15 4. X-Ray field/receptor center alignment XR8/2.17 5. Indication of field size XR8/2.18 6. X-Ray field Limitation & Alignment XR8/2.20 7. Cassette Tray/ Inspection Cleaning 8.
(NO TEXT) November 2005 ( 5 - 22) Linear MC150-C Collimator
SECTION 6.
(NO TEXT) November 2005 6.0 THEORY OF OPERATION 6.1 MECHANICAL OPERATION (6-2) Linear MC150-C Collimator The EUREKA LINEAR™ MC150-C Collimator contains two major sets of shutters, long and cross, which define the absolute X-Ray field size. There is also a fixed aperture cone which protrudes into the port of the X-Ray tube that helps reduce the effects of offfocus radiation.
with these knobs. Scales provided on a deadfront panel indicate all conventional film sizes for both 40”, 72”, 100 cm and 180 cm SID’s. All Eureka Linear Series collimators have a swivel mount configuration. Detents are located at 90 degree increments. The collimator may be oriented to any position for achieving proper X-Ray field to cassette alignment for table-top or non-bucky operation. The shutter mechanism has been precisely aligned with respect to the mounting flange at the factory.
November 2005 (6-4) Linear MC150-C Collimator
SECTION 7.
70-11236 70-04752 70-11246 70-11250 70-11248 70-11253 70-04571 70-04300 70-04572 70-04299 70-01901 70-03051 70-20024 70-10049 70-10050 70-08359 70-10122 70-10282 70-11251 70-08177 70-10810 Window - Cross Hair Knob - Front Panel Cover – Outer Wrap Cover - Front Cover - Top Cover - Rear Lamp - Light Field - DZE 24 Vac, 150W (mfg. before December 2003) Lamp – Light Field – FCS 24 Vac, 150W (mfg. after November 2003) Socket – Lamp (mfg. before December 2003) Socket – Lamp (mfg.
SECTION 8.
IR VIRC VIRL IR TRUE VCPC VCPL X-Ray Field in the Long Dimension Image Receptor (Cassette Tray) Voltage from the Cassette Sensing Element representing the Image Receptor size in the Cross Dimension Voltage from the Cassette Sensing Element representing the Image Receptor size in the Long Dimension Voltage representing the Presence of a Cassette Voltage applied to the Collimator Potentiometers in the Cross Position Voltage applied to the Collimator Potentiometers in the Long Position November 2005 (8-2)
General Electric Picker/Dunlee Machlett November 2005 B150 B160 B180 Maxiray 100 HRT, MX75 DU - 140 DU - 200 DU - 300 PX - 400 PX - 1300 PX - 1400 DX40 Series DX50 Series DX60 Series DX70 Series 2 - 11/64” 2 - 1/4” 2 - 1/4” 2 - 5/16” 2 - 1/16” 2 - 1/16” 2 - 1/16” 2 - 9/32” 2 - 5/16” 2 - 3/16” 2 - 5/16” 2 - 1/16” 2 - 3/16” 2 - 5/16” 2 - 5/16” (8-3) 17/64” 3/16” 3/16” 1/8” 3/8” 3/8” 3/8” 5/32” 1/8” 1/4” 1/8” 3/8” 1/4” 1/8” 1/8” (.637 cm) (.477 cm) (.477 cm) (.125) (.375) (.375) (.375) (.156) (.125) (.
(NO TEXT) November 2005 (8-4) Linear MC150-C Collimator
REVISIONS REV +15V 5 J3 24VAC 1 J1 1 VIN 1.0A PICO II LM317T VOUT ADJ 2 3 J1 0VAC + CR1 C1 C2 1000uF ELEC .01uF CER 237 J4 J4 J3 1 A. KREMA F1 TO 1A, J1-J5 TO J1, J3 TO J2, R5 TO 1.2K, C6 TO .22 9/25/98 A. KREMA C ADD R16, JP1, J3 7/8/99 A. KREMA D ADD J4 5/3/01 A. KREMA E ADD J1-2 CONNECTION TO J1-4 10/14/01 A. KREMA 1.2K 5% D1 1 4 7/14/98 R5 2.
Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089 Eureka BILL OF MATERIAL ITEM QTY DWG NO. SIZE 0001 0002 0003 0004 0005 0006 0007 0008 0009 0010 0011 0012 0013 0014 0015 0016 0017 0018 0019 0020 1 1 3 1 5 1 1 1 1 1 1 1 1 1 1 1 1 1 2 2 B A A A A A A A A B A A B B D D D D D D PART NO.
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089 BILL OF MATERIAL ITEM QTY DWG NO. SIZE 0021 0022 0023 0024 0025 0026 0027 0028 0029 0030 0031 0032 0033 0034 2 1 1 1 1 1 .5 .5 1 1 1 1 1 1 REF REF D A B A A A B A A A A A A A B B PART NO. 70-00078 70-02021 70-03061 70-04608 70-04748 70-04291 20-46223-2 70-06544-0 70-04701 70-04506 70-04222 70-04216 70-04217 10-04002 70-08358 70-08359 ITEM DESCRIPTION CLS Bill of Material No. 70-08359 Rev. F ECN: P0690 Vendor Ref. Part No.
Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089 Eureka BILL OF MATERIAL ITEM QTY DWG NO. SIZE 0001 0002 0003 0004 0005 0006 0007 0008 0009 0010 0011 0012 0013 0014 0015 0016 0017 0018 0019 0020 1 1 3 1 5 1 1 1 1 1 1 1 1 1 1 1 1 1 2 2 B A A A A A A A A B A A B B D D D D D D PART NO.
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089 BILL OF MATERIAL ITEM QTY DWG NO. SIZE 0021 0022 0023 0024 0025 0026 0027 0028 0029 0030 0031 0032 0033 0034 0035 36 2 1 1 1 1 1 .5 .5 1 1 1 1 1 1 1 1 REF REF D A B A A A B A A A A A A A A A B B PART NO. 70-00078 70-02021 70-03061 70-04608 70-04748 70-04291 20-46223-2 70-06544-0 70-04701 70-04506 70-04222 70-04216 70-04217 70-04758 70-04806 10-04002 70-08358 70-08359 ITEM DESCRIPTION CLS Bill of Material No. 70-08369 Rev.