EUREKA™-IP & EUERKA™-LF INFUSION PUMP OPERATOR’S MANUAL Universal Medical Technologies Inc.
This guide applies to the EUREKA-IP ™ INFUSION PUMP Catalog# 11016 This guide applies to the EUREKA-LF™ INFUSION PUMP Catalog# 11017 © 2001 Universal Medical Technologies Inc.
Contents 1.0 INTRODUCTION ................................................................................................................... 4 2.0 GENERAL DESCRIPTION ..................................................................................................... 4 2.1 PHYSICAL DESCRIPTION .................................................................................................... 4 2.1.1 PUMP DESIGN ..............................................................................................
1.0 INTRODUCTION The EUREKA™ INFUSION PUMPS are designed to provide intravenous drug therapy to patients in the hospital, home, clinic or surgery center settings. Health care clinicians (physicians, nurses, pharmacists and others) should be completely familiar with this manual before operating or dispensing this pump. The Eureka-IP and the Eureka-LF are designed and intended to be used with non emergency medications. They should be used with Antibiotic and Chemotherapy type drugs in Solution 2.
2.2 IV SOLUTION CONTAINERS The EUREKA ™ INFUSION PUMPS will accommodate Baxter 50 ml and 100 ml IV solution containers. Abbott 50 ml and 100 ml IV solution containers 50 mls may be filled to a maximum volume of 58 mls 100 mls may be filled to a maximum volume of 110 mls Typically these containers have overfill beyond their labeled volumes (typically 5%) and pharmacists must use caution when filling to be sure that additive drug/solutions not exceed the maximum volumes recommended.
2.3 INFUSION RATES AND VOLUMES 2.3.1 AVAILABLE FLOW RATES Currently the Eureka-IP Administration Set flow rates are: 50 ml/hr EUREKA ™ IV ADMINISTRATION SET 100 ml/hr EUREKA ™ IV ADMINISTRATION SET 200 ml/hr EUREKA ™ IV ADMINISTRATION SET Currently the Eureka-LF Administration Set flow rates are: 1 ml/hr EUREKA ™ IV ADMINISTRATION SET 2 ml/hr EUREKA ™ IV ADMINISTRATION SET 5 ml/hr EUREKA ™ IV ADMINISTRATION SET 2.3.
2.3.3 FLOW RATE VARIABLES The Eureka Administration Sets are designed to meet the labeled flow rates for use at room temperature of 68-75 degrees F using normal saline. Result between D5W and Normal Saline at room temperature are within normal flow rate accuracy (+/- 12%). However, increases in fluid viscosity or decreases in solution temperature can result in a decreased (slower) fluid flow rate.
Effect of Pressure (NS=Normal Saline) Pressure Effect (NS) 100ml/hr @ 0 and .65 psi 130 cc/hr (cumlative) 120 110 .65 psi 100 0 psi 90 80 70 Flow Rate Effect of Pressure (Static Back Pressure) NS 100cc @ 100cc/hr 0 PSI 0.65 PSI 15 minutes 108.6 84.4 cc/hr 30 minutes 101.8 79.1 cc/hr 45 minutes 98.4 76.3 cc/hr 60 minutes 93.2 73.7 cc/hr 60 1 9 17 25 33 41 49 57 65 73 81 89 5 13 21 29 37 45 53 61 69 77 85 93 Minutes Tubing Set flow rate is not affected by normal venous pressure.
3.0 OPERATING INSTRUCTIONS 3.1 CHANGING BATTERIES / RECHARGING / DC ADAPTER Slide open Battery Door. (Pressing on Red Circle on bottom of case will make slideing the battery door easier.) When battery door is down., disconnect battery cabling. Slide battery pack out to replace. ADAPTER Connection Charging Battery and connecting External Power Supply: Plug into connector and verify proper connection by checking External Power light is Illuminated (Green).
3.2 EXTERNAL POWER SUPPLY Utilize Eureka supplied Power Adapter ONLY. Recharge time is between 30-60 minutes during nonoperation. Charging during use may result in slightly longer times before full charge is achieved. Full charge can be determined when Battery Charged Light illuminates WHILE external power charger is connected 3.
3.
3.5 ENGAGING TUBING Follow instructions on page 11. Make sure Tubing set clamp is closed. Spike IV Bag 3.6 PLACING BAG IN PUMP Position Bag in Pump as follows: Place pump on flat surface. Place bag so that Spike Flange rests in slot before case outlet, and aligns with lever switch.
3.7 PUMP OPERATIONS 3.7.1 USING THE EUREKA ™ INFUSION PUMP 1. Engage EUREKA ™ administration set and IV solution container per administration Set instructions. 2. Turn EUREKA ™ Infusion Pump on by depressing the POWER button (1). Power Light (2) will go on. 3. Open Pump Lid- AVOID TOUCHING THE BLADDER AREA OF CAVITY CHAMBER. 4. Place Bag in pump cavity. 5. Secure EUREKA ™ IV Administration Set spike and port securely in designated area. 6. Close and latch Lid. 7. Check Battery Low light (10).
3.7.2 WARNINGS AND ALARMS 1. POWER Light GREEN (2) indicates pump is ON when illuminated. POWER up sequence at powering ON of unit consists of a continuous tone and all lights illuminated for 1 second. 2. START Light (4) will illuminate GREEN when in infusion process. Press for 1 beep to initiate. 3. STOP Light (6) will blink RED while in stop mode or pause mode is activated. STOP must be pressed for 2 beeps to prevent accidental stopages during an infusion and will emit 2 short beeps. 4.
FUNCTION KEYS START-Starts or restarts infusion cycle- Press once when Power is ON STOP- hold press (2 seconds or 2 beeps) to stop or pause operation POWER ON/OFF-Power ON (Press once) Power OFF (hold press for 2 seconds/2 beeps) Cover Release -Slide lid open latch while in STOP mode to open lid for removal or placement of IV solution container Other Safety Features Include 1. Sensor for excessive pressure (over 6.5 psi) to prevent against hazards. 2.
3.8 FLOW RATE PROFILE and TOTAL VOLUME FLOW EUREKA ™ Infusion Pump Administration Set TYPICAL FLOW PROFILE Flow Profiles Flow Rate as % of Labeled Rate 120 120 100 Flow Rate (%) 60 40 20 cc/hr 80 Total cc's % of Labeled Rate 100 80 Total cc's 60 cc/hr 40 20 0 0 1 3 2 5 4 7 6 9 8 Time Intervals 11 10 13 12 5 15 10 25 20 35 30 45 40 55 50 65 60 70 Time (Minutes) TYPICAL FLOW RATE RESULTS FOR ADMINISTRATION SET Labeled Set Rate 100 ml/hr Max.
3.9 DISPOSABLE Instructions (Typical) Eureka ™ IV Administration Set Disposable Delivery accuracy : +/- 12% based on 100 ccs Normal Saline Viaflex container Fluid pathway is sterile and nonpyrogenic. Priming Volume is 1.2 ccs For intravenous antibiotic delivery only Caution: If end protector is missing or not in place then the sterility of this device is not guaranteed. DO NOT use this device if end protector is either missing or not in place.
To Disconnect: 1. At pump signal of Check Status verify complete infusion. a. If infusion has not completed then check the IV tubing line for crimping , closed clamps or obstructed tubing. b. If infusion the time of infusion is finished then proceed to next step. 2. Clamp close tubing by closing clamp on tubing. 3. Disconnect from IV Site. 4. Remove bag and tubing from EUREKA ™ INFUSION PUMP. 5. Dispose of IV bag and administration set in appropriate manner.
4.0 REFERENCE SECTION 4.1 GLOSSARY Administration Set- Tubing and spike attachment that connects solution container and patient’s access device Aspirate - To allow air to vent off. (Also see Purge) Battery Charged - Battery has reached comletion of charge cycle. NOTE: A fully charged battery may take upto 10 minutes before the indicator signals that the charge cycle is complete.
4.2 PUMP CLEANING & MAINTENANCE Use soft damp cloth only. Cloth may be dampened with one of the following solutions Soap and water 70% Ethyl Alcohol 70% Chlorhexidine in Alcohol 70% Isopropyl Alcohol DO NOT immerse pump in cleaning fluid or water AVOID contact with bladder membrane DO NOT UN-CAP needles until lid to Pump is closed. This is to prevent a bladder puncture from needle sticks. 4.3 PUMP EXPOSURE TO RADIATION OR MAGNETIC RESONANCE IMAGING (MRI) Pump should not be subject to direct irradiation.
4.4 SPECIFICATIONS Model - IP Catalog Number 11016 Eureka-IP Pump with Accessories (US) Model - LF Catalog Number 11017 Eureka-LF Pump with Accessories (US) Volume range 50-58 ml. for 50 ml. IV solution Container 100-110 ml. for 100 ml.
4.6 LIMITED WARRANTY INFORMATION Universal Medical Technologies (UMT) warrants to the original purchaser that the Infusion Pump (not including accessories) is free from defects in materials and workmanship under proper use, if used in accordance with this manual, for a period of 90 days from the date of sale to original purchaser. An optional warranty extending this coverage till one year may be purchased at time of sale.