Specifications

-3.4-
SYS 1 SYS 2 SYS 3 DCU OK Code Name
OFF OFF RED RED
130
PRC_STATUS_DATAIO_FAILURE_INDICATION
SYS 1 SYS 2 SYS 3 DCU OK Code Name
RED RED OFF RED
042
PRC_STATUS_GRABBER_FAILURE_INDICATION
SYS 1 SYS 2 SYS 3 DCU OK Code Name
RED YELLOW RED RED
186
PRC_STATUS_GRABBER_IMAGE_TOO_BIG
SYS 1 SYS 2 SYS 3 DCU OK Code Name
RED OFF RED RED
138
PRC_STATUS_CONFIG_FAILURE_INDICATION
SYS 1 SYS 2 SYS 3 DCU OK Code Name
RED GREEN RED RED
154
PRC_STATUS_CONFIG_INVALID_START_TAG
SYS 1 SYS 2 SYS 3 DCU OK Code Name
GREEN GREEN RED RED
150
PRC_STATUS_CONFIG_INVALID_SIZE
SYS 1 SYS 2 SYS 3 DCU OK Code Name
RED GREEN GREEN RED
090
PRC_STATUS_CONFIG_INVALID_FLASH_HANDLE
Description:
Problem with ANALOG/DIGITAL interface board inside DPU.
Selftest program run during boot should indicate which board.
Action:
Replace faulty board.
Description:
One of the active/enabled video-channels can't grab from the channel. This can be
either because radar is turned OFF or wrong configuration.
Action:
If radar is turned OFF, press ACK on RAP to stop alarm. Error will clear when radar
is turned ON again. If Radar is ON, check installation(cables) and configuration.
Description:
Video image captured from radar is large than expected.
Action:
Check quality of captured image, colormask(video-channels) and re-adjust
framegrabber.
Description:
Unexpected error when trying to read CONFIG.
Action:
Check that installed firmware version is correct for connected DRU. If this is a new
DRU, CONFIG_area needs to be 'Cleared'.
Description:
CONFIG_area in the DRU is clean. This should only happen during
installation/service. This will also be displayed after first boot after clearing
CONFIG_area. It will flash every 15 sec with SYS1/SYS2/SYS3 off and this code.
Action:
Restart VR-5000.
Description:
Size of CONFIG_area in the DRU is different from expected one. This should only
happen during installation/service and upgrade of the system software.
Action:
Clear CONFIG_area and restart VR-5000. This will install a ’default’ minimum
configuration.
Description:
CONFIG_area in the DRU is corrupt/wrong/un-initialized. This should only happen
during installation/service and if using a DRU not initialized by the VR-5000 it is
connected to.
Action:
Clear CONFIG_area and restart VR-5000. This will install a ’default’ minimum
configuration.