User's Manual

22
This equipment complies with FCC/IC/EU RF radiation exposure limits set forth for an uncontrolled
environment. The highest SAR value for the MD-150 device is <1.562W/kg.
In order to maintain compliance as a Class 1 Medical device, the switches that are used in combination
with the MD-150 device must fulfill the isolation requirements of IEC/EN 60601-1 based on 15 VDC.
They must be electrically floating switches (switches that are not grounded).
B2 CE Statement
This equipment complies with the requirements relating to electromagnetic compatibility, the essential
protection requirement of Electromagnetic Compatibility (EMC) Directive 2014/30/EU on the
approximation of the laws of the Member States relating to electromagnetic compatibility and Radio
Equipment Directive (RED) 2014/53/EU to meet the regulation of the radio equipment and
telecommunications terminal equipment.
B3 Directives and Standards
The MD-150 complies with the following directives:
Medical Device Regulation (MDR)
Low voltage Directive 2014/35/EU
Electromagnetic Compatibility (EMC) Directive 2014/30/EU
Radio Equipment Directive (RED) 2014/53/EU
RoHS2 Directive 2011/65/EU
WEEE Directive 2012/19/EU
Reach Directive 2006/121/EC, 1907/2006/EC Annex 17
Batteries Directive 2013/56/EU
The device has been tested to comply with FCC, IEC 60601-1, IATA/IEC62133, and other relevant
standards for the intended markets.