Specifications
SPECIFICATIONS
SAFETY DEVICES
ACCESSORIES
●
Leakage breaker
●
Electrical compartment door switch
●
Door switch
●
Thermal fuse
●
Temperature switch for air circulator
●
Heater wiring breaker
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
●
Shelf (stainless steel wire) 2
(stainless steel plate for type 332)
●
Shelf bracket (stainless steel) 2 sets (4)
●
Cartridge fuse 2
●
User's manual 1 set
8
Shelf pitch, quantity and load resistance
Model PV
−
212 PV
−
222 PV
−
232 PV
−
332 PVH
−
212 PVH
−
222 PVH
−
232 PVH
−
332
System Forced hot-air circulation / ventilation system
Performance
*1
Temperature range
*2
Ambient temp.
+
20
℃
(
+
68°F) to
+
200
℃
(
+
392°F) Ambient temp.
+
20
℃
(
+
68°F) to
+
300
℃
(
+
572°F)
Temperature
uctuation
*2
±
0.2
℃
at
+
100
℃
(
+
212°F),
+
200
℃
(
+
392°F)
±
0.2
℃
at
+
100
℃
(
+
212°F),
+
200
℃
(
+
392°F),
±
0.3
℃
at
+
300℃ (
+
572°F)
Temperature
uniformity
*2
±
1.0
℃
at
+
100 (
+
212°F),
±
2.0
℃
at
+
200
℃
(
+
392°F)
±
1.0
℃
at
+
100
℃
(
+
212°F),
±
2.0
℃
at
+
200
℃
(
+
392°F),
±
3.0
℃
at
+
300℃ (
+
572°F)
Temperature
heat-up time
Ambient temp. to
+
200
℃
(
+
392°F)
within 40 min.
Ambient temp. to
+
300
℃
(
+
572°F)
within 60 min.
Construction
Exterior material Cold rolled rust-proof steel plate, Melamine resin coating
Interior material Stainless steel plate
Insulation material Glass wool
Heater Sheathed heater
Air circulator Stainless steel sirocco fan
Damper Circulation
/
Ventilation (manual switching
)
Fittings
Power cable (approx 2m from chamber), Specimen power supply control terminals
(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A)
Inside dimensions
W
×
H
×
Dmm (in)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
600
×
900
×
600
(23.6
×
35.4
×
23.6)
600
×
1200
×
600
(23.6
×
47.2
×
23.6)
800
×
1200
×
800
(31.5
×
47.2
×
31.5)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
600
×
900
×
600
(23.6
×
35.4
×
23.6)
600
×
1200
×
600
(23.6
×
47.2
×
23.6)
800
×
1200
×
800
(31.5
×
47.2
×
31.5)
Outside dimensions
*3
W
×
H
×
Dmm (in)
770
×
1200
×
925
(30.3
×
47.2
×
36.4)
770
×
1500
×
925
(30.3
×
59
×
36.4)
770
×
1800
×
925
(30.3
×
70.9
×
36.4)
1030
×
1800
×
1145
(40.6
×
70.8
×
45.1)
770
×
1200
×
925
(30.3
×
47.2
×
36.4)
770
×
1500
×
925
(30.3
×
59
×
36.4)
770
×
1800
×
925
(30.3
×
70.9
×
36.4)
1030
×
1800
×
1145
(40.6
×
70.8
×
45.1)
Capacity (L) 216 324 432 768 216 324 432 768
Weight (kg) 165 190 210 325 165 190 210 325
Allowable ambient conditions
Temperature: 0 to
+
40
℃
(
+
32 to
+
104°F) Humidity: to 75
%
rh
Utility requirements
Power supply
(Voltage uctuation:
±
10
%
of rated value)
200
/
220
/
230
/
240V AC
1
φ
50
/
60Hz
200
/
220V AC
3
φ
3W 50
/
60Hz
200
/
220
/
230
/
240V AC
1
φ
50
/
60Hz
200
/
220V AC 3
φ
3W 50
/
60Hz
─
380V AC
3
φ
4W 50Hz
─
380V AC
3
φ
4W 50Hz
Max. power consumption
(kVA)
4.0 4.8 5.8 6.8 4.0 5.8 6.2 8.8
*1 Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
Model
Shelf
pitch
Shelves
Shelf load
resistance
*1 *2
Chamber total
load resistance
*1
PV(H)-212
50mm
11
25kg
200kg
PV(H)-222 17
PV(H)-232 23
PV(H)-332 80mm 14 45kg
*1 Including shelf weight
*2 Equally distributed load