Specifications

SPECIFICATIONS
SAFETY DEVICES
ACCESSORIES
Leakage breaker
Electrical compartment door switch
Door switch
Thermal fuse
Temperature switch for air circulator
Heater wiring breaker
Upper and lower temperature limit alarm
(built inside temperature controller)
Overheat protector
Cartridge fuse
Specimen power supply control terminal
Shelf (stainless steel wire) 2
(stainless steel plate for type 332)
Shelf bracket (stainless steel) 2 sets (4)
Cartridge fuse 2
User's manual 1 set
8
Shelf pitch, quantity and load resistance
Model PV
212 PV
222 PV
232 PV
332 PVH
212 PVH
222 PVH
232 PVH
332
System Forced hot-air circulation / ventilation system
Performance
*1
Temperature range
*2
Ambient temp.
20
(
68°F) to
200
(
392°F) Ambient temp.
20
(
68°F) to
300
(
572°F)
Temperature
uctuation
*2
±
0.2
at
100
(
212°F),
200
(
392°F)
±
0.2
at
100
(
212°F),
200
(
392°F),
±
0.3
at
300℃ (
572°F)
Temperature
uniformity
*2
±
1.0
at
100 (
212°F),
±
2.0
at
200
(
392°F)
±
1.0
at
100
(
212°F),
±
2.0
at
200
(
392°F),
±
3.0
at
300℃ (
572°F)
Temperature
heat-up time
Ambient temp. to
200
(
392°F)
within 40 min.
Ambient temp. to
300
(
572°F)
within 60 min.
Construction
Exterior material Cold rolled rust-proof steel plate, Melamine resin coating
Interior material Stainless steel plate
Insulation material Glass wool
Heater Sheathed heater
Air circulator Stainless steel sirocco fan
Damper Circulation
/
Ventilation (manual switching
Fittings
Power cable (approx 2m from chamber), Specimen power supply control terminals
(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A)
Inside dimensions
 
W
×
H
×
Dmm (in)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
600
×
900
×
600
(23.6
×
35.4
×
23.6)
600
×
1200
×
600
(23.6
×
47.2
×
23.6)
800
×
1200
×
800
(31.5
×
47.2
×
31.5)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
600
×
900
×
600
(23.6
×
35.4
×
23.6)
600
×
1200
×
600
(23.6
×
47.2
×
23.6)
800
×
1200
×
800
(31.5
×
47.2
×
31.5)
Outside dimensions
*3
W
×
H
×
Dmm (in)
770
×
1200
×
925
(30.3
×
47.2
×
36.4)
770
×
1500
×
925
(30.3
×
59
×
36.4)
770
×
1800
×
925
(30.3
×
70.9
×
36.4)
1030
×
1800
×
1145
(40.6
×
70.8
×
45.1)
770
×
1200
×
925
(30.3
×
47.2
×
36.4)
770
×
1500
×
925
(30.3
×
59
×
36.4)
770
×
1800
×
925
(30.3
×
70.9
×
36.4)
1030
×
1800
×
1145
(40.6
×
70.8
×
45.1)
Capacity (L) 216 324 432 768 216 324 432 768
Weight (kg) 165 190 210 325 165 190 210 325
Allowable ambient conditions
Temperature: 0 to
40
(
32 to
104°F) Humidity: to 75
rh
Utility requirements
Power supply
(Voltage uctuation:
±
10
of rated value)
200
/
220
/
230
/
240V AC
1
φ
50
/
60Hz
200
/
220V AC
3
φ
3W 50
/
60Hz
200
/
220
/
230
/
240V AC
1
φ
50
/
60Hz
200
/
220V AC 3
φ
3W 50
/
60Hz
380V AC
3
φ
4W 50Hz
380V AC
3
φ
4W 50Hz
Max. power consumption
(kVA)
4.0 4.8 5.8 6.8 4.0 5.8 6.2 8.8
*1 Values assume circulatory operation with no specimens at an ambient temperature of
23
℃±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
Model
Shelf
pitch
Shelves
Shelf load
resistance
*1 *2
Chamber total
load resistance
*1
PV(H)-212
50mm
11
25kg
200kg
PV(H)-222 17
PV(H)-232 23
PV(H)-332 80mm 14 45kg
*1 Including shelf weight
*2 Equally distributed load