Specifications
Suitable for heat deterioration test
Based on the PH Temperature Chambers, these
models incorporate a detachable rotating specimen
rack and is especially designed for heat deterioration
testing of rubbers and plastics including polyesters
and vinyls.
Equipped with a rotating specimen rack
The rack drive unit is installed inside, enhancing
function and lending them a simple appearance.
By removing the rack, this equipment may also be
operated as an ordinary temperature chamber.
●
Shelf (stainless steel wire) 2
●
Shelf bracket (stainless steel) 2 sets (4)
●
Cartridge fuse 2
●
Specimen clip type102 50
type202 100
●
Shaftinsulationfilters
1 set
●
User's manual 1 set
●
Leakage breaker
●
Electrical compartment door switch
●
Thermal fuse
●
Temperature switch for air circulator
●
Heaterwiringbreaker
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
Test area
SAFETY DEVICES
SPECIFICATIONS
ACCESSORIES
TEMPERATURE CHAMBER WITH ROTATING SPECIMEN RACK
GPH
(
H
)
+ 200℃/+ 300℃
16
Model
GPH−102 GPH−202 GPHH−102 GPHH−202
System
Forced hot-air circulation / ventilation system
Performance
*1
Temp.range
*2
Ambient temp.
+
20
℃
(
+
68°F)
to
+
200
℃
(
+
392°F)
Ambient temp.
+
20
℃
(
+
68°F)
to
+
300
℃
(
+
572°F)
Temp. uctuation
*2
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
±
0.2
℃
at
+
300
℃
(
+
572°F)
Temp.uniformity
*2
±
0.5
℃
at
+
100
℃
(
+
212°F)
±
1.5
℃
at
+
200
℃
(
+
392°F)
±
0.5
℃
at
+
100
℃
(
+
212°F)
±
1.5
℃
at
+
200
℃
(
+
392°F)
±
2.5
℃
at
+
300
℃
(
+
572°F)
Temp.heat-up time
Ambient temp.
to
+
200
℃
(
+
392°F)
within 40min.
Ambient temp.
to
+
300
℃
(
+
572°F)
within 60min.
Specimen rack rotating unit
Number of racks
1 2 1 2
Outside diameter
320mm (12.6in.)
Available numbers
of specimens/weight
56pcs per rack (up to 0.7kg)
Specimen clip
50pcs per rack
RPM of specimen rack
5rpm
/
50Hz, 6rpm
/
60Hz
Motor
1φ 15W
Viewing
window
W
×
Hmm (in)
190
×
340(7.48
×
13.39)
————
Construction
Heat-resistant
reinforced glass
3-plate sets
————
Chamber lamp
5.5W
incandescent lamp
————
Fittings
Power cable (approx 2m from chamber),
Specimen power supply control terminals
(relay contact is opened during malfunction.
Voltage capacity 250V AC 3A)
Inside dimensions
W
×
H
×
Dmm (in)
450
×
450
×
450
(17.7
×
17.7
×
17.7)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
450
×
450
×
450
(17.7
×
17.7
×
17.7)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
Outside dimensions
*3
W
×
H
×
Dmm (in)
1040
×
820
×
635
(91
×
32.3
×
25
)
1190
×
970
×
785
(46.9
×
38.2
×
30.9)
1040
×
820
×
635
(91
×
32.3
×
25)
1190
×
970
×
785
(46.9
×
38.2
×
30.9)
Capacity (L) 91 216 91 216
Weight (kg) 95 130 95 130
Allowable ambient conditions
Temp.: 0 to
+
40
℃
(
+
32 to
+
104°F) Humid.: to 75
%
rh
Utility
requirements
Power supply
(
±
10
%
of rated value)
200
/
220
/
230
/
240V AC 1
φ
50
/
60Hz
Max. power consumption
2.0 kVA 2.7 kVA 3.8 kVA
*1
Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.