Specifications
Low oxygen level testing
Equipped with a non-oxidizing gas intake structure
which fills the chamber with non-oxidizing gas such
as CO
2
or N
2
for heat treatment or temperature
characteristics testing requiring low oxygen
concentration atmosphere.
Hermetically sealed configuration
The chamber is hermetically sealed to decrease oxygen
inside the chamber. The inner stainless steel plate is
seamless welded with argon gas.
O
2
concentration indicator controller (optional)
An optional O
2
concentration indicator controller
equipped with an oxygen sensor is available. It
allows precise regulation of the O
2
level throughout
the range 0.5 to 21% (using N
2
).
●
Shelf (stainless steel wire) 2
●
Shelf bracket (stainless steel) 2 sets (4)
●
Cartridge fuse 2
●
User's manual 1 set
●
Leakage breaker
●
Electrical compartment door switch
●
Thermal fuse
●
Temperature switch for air circulator
●
Heater wiring breaker
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
ANAEROBIC TEMPERATURE CHAMBER
IPH
(
H
)
+ 200℃/+ 300℃
SPECIFICATIONS
ACCESSORIES
SAFETY DEVICES
15
Model IPH
−
202 IPHH
−
202
System Forced hot-air circulation system
Performance
*1
Temp. range
*2
Ambient temp.
+
20
℃
(
+
68°F)
to
+
200
℃
(
+
392°F)
Ambient temp.
+
20
℃
(
+
68°F)
to
+
300
℃
(
+
572°F)
Temp.
uctuation
*2
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
±
0.2
℃
at
+
300
℃
(
+
572°F)
Temp.
uniformity
*2
±
0.5
℃
at
+
100
℃
(
+
212°F)
±
1.5
℃
at
+
200
℃
(
+
392°F)
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
±
0.2
℃
at
+
300
℃
(
+
572°F)
Temp.
heat-up time
Ambient temp. to
+
200
℃
(
+
392°F) within 40min.
Ambient temp. to
+
300
℃
(
+
572°F) within 60min.
Gas intake unit
Fluid
CO
2
, N
2
gas (ordinary temperature, dry gas)
Fluid pressure
Allowed max. pressure: 2.0MPa (Gauge) (primary side of valve)
Secondary side is adjusted with the valve to 0.05MPa (Gauge)
Flow rate
Max. ow rate: 20 L
/
min. (0.05MPa (Gauge), 20
℃
)
Chamber O
2
level
0.5
%(
lowest
)
Chamber injector pressure
29Pa (Gauge) and over (at max ow rate)
Valve
1/4" brass needle valve
Pressure gauge
φ
75mm embedded type class 2.5
Scale range: 0
〜
0.1MPa (Gauge)
Flow meter
Floating type (provided with needle valve for ow rate control)
Scale range
0 to 30L
/
min. N
2
gas
Safety valve
Trip pressure: 2.0kPa (Gauge)
Gas inlet
1
/
4" ring joint
Fittings
Power cable (approx 2m from chamber),
Specimen power supply control terminal
Inside dimensions (in)
W600mm
×H
600mm
×D
600mm (23.6
×
23.6
×
23.6)
Outside dimensions (in)
*3
W119 0 mm
×H
970mm
×D
785mm (46.9
×
38.2
×
30.9)
Capacity (L)
216
Weight (kg)
130
Allowable ambient conditions Temp.: 0 to
+
40
℃
(
+
32 to
+
104°F) Humid.: to 75
%
rh
Utility
requirements
Power supply
(
±
10
%
of rated value)
200
/
220
/
230
/
240V AC 1
φ
50
/
60Hz
Max. power consumption
2.7 kVA 3.8 kVA
*1
Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.