Specifications

Tabbed user interface
Provides high resolution and fast display.
Tabs at the bottom of the screen make it
easy to activate any section.
Eco operation settings are easier than ever.
Register test patterns
Up to 40 patterns for program operation
and 3 patterns for constant operation can
be registered.
Trend graph output
Trend graphs can be displayed according
to set conditions and data can be recorded
in internal memory. Data can also be
recorded directly onto USB memory for
backup purposes. Measurement intervals
and other settings can be changed.
Information screen
The infor mation section displays
information on the chamber status, such
as defrosting, or humidifying tray water
auto refill.
Messages displayed according to operating
timer inform you of the humidifier
maintenance checks, condenser filter or
strainer checks.
It is also possible to register and use
additional notification items.
Multilingual display
A simple operation changes display text
to Japanese, Chinese (simplif ied,
traditional), or Korean. Select the language
that suits your needs.
High-speed processing N-Instrumentation features improved operability
and legibility
Characteristics
N-Instrumentation
Operating mode Constant operation, program operation, remote operation, stop
Setting range
Constant setup 3 patterns
Program setup 40 patterns (99 steps per program)
Language English, Japanese, Chinese, Korean (switch without restarting)
External
memory
function
• Interface
USB 2.0 standard compliant (A-type connector)
• Supported functions
Write log, Read/Write program (application software: Pattern
Manager Lite), write backtrace
Web function
• Interface
Ethernet port (100base-TX)
• Server functions
Remote monitor/setup (constant, program)/operation, email alert,
Web Manager functions
• Browser
Windows internet Explorer 7 or 8
N-Instrumentation
6