Specifications

CLEAN TEMPERATURE & HUMIDITY CHAMBER
PCR
−20
to
100℃・30
to
90%rh
Model PCR3J
System Balanced Temperature and Humidity Control system (BTHC system)
Performance*
1
Temp. & humidity range
−20 to +100°C/30 to 90% rh
(lowest attainable temperature in an ambient temperature of 0 to +30°C)
Refer to diagram of temperature & humidity controllable range on page 26.
Temp. & humidity fluctuation ±0.5°C2.5% rh
Temperature gradient 5.0°C
Temperature variation in space 5.0°C
Temperature rate of change
Heat up rate: 1.5°C/min.
Pull down rate: 1.0°C/min.
Temperature extremes
achievement time
Heat up time: from +20 to +10C 55 min.
Pull down time: from +20 to −20°C 45 min.
Cleanliness*
2
Class5 (Particle diameter: 0.5μm)
Construction
Exterior material Stainless steel plate: 18 Cr stainless steel plate, hairline finish
Test area material Stainless steel plate: 188 Cr−Ni stainless steel plate, 2B polish
Heater Nichrome strip wire heater
Humidifier 18-12−2.5 Cr−Ni−Mo stainless steel sheathed heater (surface evaporating system)
Cooler (dehumidifier) Plate fin cooler (Doubles as dehumidifier)
HEPA filter Collection efficiency: 99.97% or higher in 0.m single distribution D.O.P. test
Air circulator Sirocco fan
Water
supply
Supply system Pump out system
Water tank 16 L
Refrig-
eration
unit
System Mechanical type single-stage compression cooling
Compressor
Rotary compressor (R404A)
Reciprocating compressor (R404A)
Refrigerator capacity 1.5 kW + 0.4 kW
Expansion mechanism Electronic expansion valve, capillary tube
Required exhaust equipment Exhaust flow rate: 16m
3
/ min. (50Hz);18m
3
/min. (60Hz); Chamber connection port: ø123mm
Capacity 312 L
Chamber total load resistance 100 kg
Dimensions*
3
Inside dimensions
(W x H x D mm)
600 x 650 x 800
Outside dimensions
(W x H x D mm)
1010 x 1880 x 1273
Weight 445 kg
Utility requirements
Allowable ambient conditions +5 to +35°C/up to 75% rh
Power
supply
200V AC 3ø 50/60 Hz 23.5 A
220V AC 3ø 60 Hz*
4
22.0 A
380V AC 3ø 50 Hz*
4
11.0 A
400V AC 3ø 50 Hz*
4
10.5 A
*1 The performance values are based on IEC60068-3-5:2001 and IEC60068-3-6:2001;
Performance figures are given for a +23°C ambient temperature, relative humidity of 65±20%rh, rated voltage, and no specimen inside the test area.
*2 When temperature is stable, the cleanliness is according to JIS B9920:2002 (equivalent to FED-STD-209D Class 100).
The Class 5 cleanliness cannot be maintained when the door is open. Do not open the door when operating at temperatures below 0°C
*3 Excluding protrusions. Dimension indicated in ( ) includes protrusion.
*4 Compliance with CE Marking