Specifications

INSTALLATION REQUIREMENTS
Model PR PL PSL PHP PDR PDL
PCR
PU PG
Humidifier water
supply
Use pure water with a conductivity of 0.1 to 10 μS/cm supplied from the tank.
Drainage
Drain holes are positioned at the bottom of the rear panel (150 mm above the floor).
Prepare 1 drain hose for temperature and humidity use and 1 drain hose for
continuous water supply use (option).
Hose outer diameter: 18 mm, inner diameter: 12 mm
Length: approximately 1 m
Installation
space
B
C
A A
17
Model
PR
PL
PU PSL
PG PHP PDR
PDL PCR
Type 1 Type 2 Type 3 Type 4 Type 2 Type 4 Type 2 Type 3 Type 4 Type 3 Type 4 Type 3
Side: A
Space to manipulate the cable port and adjuster feet, to connect the power supply
and the water supply and drain pipes, and to perform maintenance is required.
(We recommend 30 cm or more.)
Front: B
(cm)
70 80 120 80 120 70 80 120 80 120 80
Rear: C
Space to pass the water drain hose through and to perform maintenance in is required.
(We recommend 60 cm or more.)
This is not required if the chamber will be pulled out when maintenance is performed.
Top 60 cm or more