Platinous J Series Temperature & Humidity Chamber · Low Temperature (& Humidity) Chamber Ultra Low Temperature (& Humidity) Chamber · High Temperature & Humidity Chamber Low Humidity Type (Low) Temperature & Humidity Chamber · Clean Temperature & Humidity Chamber CAT.NO.
The Platinous J Series – The next generation of environmental test chambers Its rich collection of advanced features has quickly made the Platinous J Series the environmental test chamber of choice over the world. At the very top of the list of impressive Platinous J Series features is a dramatic reduction in power consumption, thanks to a new energy saving, highly reliable cooling system. The new N-instrumentation allows J Series chambers to support easy-to-use networking system.
Contents ● Communication Network p.3 to 5 ● The N Controller p.6 to 7 ● Energy Saving Features p.8 to 9 ● Common Features p.10 to 13 ● High Temperature & Humidity Chamber PHP p.14 ● Clean Temperature & Humidity Chamber PCR p.15 ● Low Humidity Type (Low) Temperature & Humidity Chamber PDR/PDL p.15 ● Series Configuration p.16 ● Installation Requirements p.17 ● Specifications p.18 to 25 ● Temperature & Humidity Control Range Chart Type 3 p.26 ● Option list by model p.
Characteristics Network Featured Web Manager for remote monitoring and operation Remote monitoring and control (Ethernet connection) ● Email alert LAN PC Intranet (company LAN) Mail server Platinous J Mobile phone Smartphone A unique web application allows the u s e r t o mo n it o r t he ch a mb e r, s e t programs, and start and stop operation from a PC connected to the chamber PC LAN’s port.
Characteristics Network Programming (test profile) ● Programming ● Run/Stop ● Chamber monitor ● E-mail alert Program patterns stored on the chamber can be modified from the web browser. Run/Stop Even when the chamber is OFF (but the breaker is ON), it is possible to select the desired program pattern and start chamber operation. It is also possible to turn the chamber power OFF from the PC. Recording and management of data The Web Manager saves data (approx.
Characteristics USB Pattern Manager Lite software: Get the most out of USB memory Replicate program patterns ● Program copy and computer editing C opy p r og r a m p at t e r n s f r om one chamber to transfer to another without using a PC. * USB memory is not included. Copy Pattern Manager Lite software All chambers are delivered with the Pat ter n Manager Lite, computer application software used to edit program patterns, display graphs of temperature and humidity and log data.
Characteristics N-Instrumentation High-speed processing N-Instrumentation features improved operability and legibility Tabbed user interface Provides high resolution and fast display. Tabs at the bottom of the screen make it easy to activate any section. Eco operation settings are easier than ever. Register test patterns Up to 40 patterns for program operation and 3 patterns for constant operation can be registered.
Characteristics N-Instrumentation ● Detailed settings for each step ● Constant setup Up to 40 patterns for program operation can be registered. Configure settings such as refrigerator and time signal settings. Up to 3 patterns for constant operation can be registered. ● Sampling settings ● Trend graph display ● Graph data writing settings Select the recording cycle. 1-second interval: 18 hours 5-minute interval: 227 days The actual measured values and the set values are displayed on a graph.
Characteristics Energy savings Energy savings and high performance achieved thanks to the refrigeration system • Temperature control Operating conditions 0.50 1.00 1.50 2.00 2.50 PL-3 +60°C PR-3 Energy savings rate PU-3 66% max. PL-3 +20°C PR-3 PU-3 • Temperature & humidity control Operating conditions +85°C/ 85%rh +20°C/ 40%rh 0.50 1.00 Power consumption (kWh) 2.00 2.50 1.50 Energy savings rate PL-3 70% max.
Characteristics On chambers equipped with the 400 W refrigerator, the Smart Refrigerator & Dehumidifier System controls the operation when stable at constant ranges above 50°C /40%rh, after the chamber activated the sub refrigerator to run at minimum capacity. Chambers using this double energy saving control can run with the best energy-saving rates. For example, the PL−3J power consumption can be cut by 70% max. under 85°C /85%rh conditions.
Characteristics PR · PL· PSL 100 98 90 90 TheRelative refrigeration system features an electronic 80 humidity Standard control temperature 70 auto-expansion valve with non-step % rh and humidity 60 control range which realizes high-precision temperature 55 50 and humidity control within a wide low 40 temperature 32 area of +10ºC and low humidity 30 Low humidity area of 20%rh 20 (at +70 to +85ºC).
Characteristics All-glass wide-view door (Option) PR-2·3·4 PL-2·3·4 PU-2·3·4 An all-glass wide-view door provides an unrestricted view of every bit of space inside the chamber. This option is the optimum choice when observation of or operations on a specimen are required during testing. The entire surface of the door is made of multi-layer EC (metal film deposition) glass that withstands temperatures from -40 to +120ºC (+100ºC when hand-in ports are equipped).
Characteristics ※当社器比 Automatic humidifier water replacement Humidifier stagnant water contains impurities and is a cause of trouble, so the chamber now features a function that automatically replaces the water at the period set from the controller screen. Chamber detailed settings Water supply setting Water supply system Several options to supply water to the chamber are offered, including direct tap water connection, pure water, additional tanks, etc.
Characteristics Backtrace function When the chamber stops because of trouble, the operation state just before the chamber stops is automatically recorded and saved. Saved data can be transferred by USB memory. Attach this data file to an email to ESPEC, and we will perform troubleshooting. Online diagnostics services Diagnostics service is available using the backtrace data from the time of trouble.
Characteristics PHP High temperature & humidity chamber featuring a heat pipe for cooling 消費電力量比較(例) Jシリーズ Kシリーズ ・ 温度制御 消費電力量 (kWh) 運転条件 0.50 1.00 1.50 2.00 2.50 PL-3 +60℃ PR-3 66%ダウン 省エネ率 最大 % 66 PU-3 56%ダウン Test 55%ダウ ン can be conducted at 95°C/95%rh 59%ダウン PL-3 +20℃ PR-3 High Temperature & High Humidity Chamber (Type 2, Type 3, Type 4) PU-3 Structure diagram of heat pipe Low-temperature area ・ 温湿度制御 0.50 運転条件 1.00 1.
Characteristics PCR·PDR·PDL ISO Class 5 Clean temperature & humidity chamber The clean temperature & humidity chamber employs a HEPA filter to realize ISO Class 5 cleanliness in humidity control. Low humidity type temperature chamber with expanded low humidity range 5 to 98%rh With the independently-developed rotary regenerative dehumidifier method, the low humidity range is expanded to 5%rh (at +60ºC). The humidity range in low-temperature range (+10 ºC/15%rh) can also be controlled.
Temperature Chambers Temperature & Humidity Chambers SERIES Model Temperature PR Temperature & Humidity Chamber −20 to +100°C −20 to +150°C −20 to +180°C PL Low Temperature & Humidity Chamber −40 to +100°C −40 to +150°C −40 to +180°C PSL Ultra Low Temperature & Humidity Chamber −70 to +100°C −70 to +150°C −70 to +180°C* PHP High Temperature & Humidity Chamber ambient temperature +10 to +100°C PDR Low Humidity Type Temperature & Humidity Chamber −20 to +100°C Humidity 20 to 98%rh Inside capac
INSTALLATION REQUIREMENTS Model Humidifier water supply PR PL PSL PHP PDR PDL PCR PU PG ─ Use pure water with a conductivity of 0.1 to 10 μS/cm supplied from the tank. Drain holes are positioned at the bottom of the rear panel (150 mm above the floor). Prepare 1 drain hose for temperature and humidity use and 1 drain hose for continuous water supply use (option).
PR −20 to +100℃(+150℃/+180℃) ・20 to 98%rh TEMPERATURE & HUMIDITY CHAMBER Model PR−1J System Temp. & humidity fluctuation Temp. gradient 3.0°C Temperature variation in space 1.5°C Heat up time: from +20 to +100°C 30 min. Pull down time: from +20 to −20°C 40 min. 800 W Allowable heat load*2 Exterior material 1100 W Stainless steel plate: 18−8 Cr−Ni stainless steel plate, 2B polish Heater Nichrome strip wire heater Humidifier 18-12−2.
PL −40 to +100℃(+150℃/+180℃) ・20 to 98%rh LOW TEMPERATURE & HUMIDITY CHAMBER Model PL−1J System Temp. & humidity fluctuation Temperature gradient 3.0°C Temperature variation in space 1.5°C Heat up rate: 3.0°C/min. Pull down rate: 2.0°C/min. Temperature rate of change Heat up time: from +20 to +100°C 30 min. Pull down time: from +20 to −40°C 45 min. 850 W Allowable heat load*2 Exterior material 1400 W Nichrome strip wire heater Humidifier 18−12−2.
PSL −70 to +100℃(+150℃/+180℃) ・20 to 98%rh ULTRA LOW TEMPERATURE & HUMIDITY CHAMBER Model PSL-2J System −70 to +100°C/20 to 98% rh (lowest attainable temperature in an ambient temperature of 0 to +30°C) Refer to diagram of temperature & humidity controllable range on page 26. Temp. & humidity range Performance*1 PSL-4J Balanced Temperature and Humidity Control system (BTHC system) Temp. & humidity fluctuation ±0.3°C/±2.5% rh Temperature gradient 3.
PHP Ambient temperature +10 to +100℃・40 to 98%rh HIGH TEMPERATURE & HUMIDITY CHAMBER Model PHP−2J Performance*1 System Temp. & humidity fluctuation ±0.3°C/±2.5% rh Temperature gradient 3.0°C Temperature variation in space 1.5°C 300 W load*2 Exterior material Stainless steel plate: 18−8 Cr−Ni stainless steel plate, 2B polish Heater Nichrome strip wire heater Humidifier 18−12−2.
PDR · PDL 5 to 98%rh・ −20 to +100℃/−40 to +100℃ LOW HUMIDITY TYPE (LOW) TEMPERATURE & HUMIDITY CHAMBER Model PDR−3J System Temp. & humidity range Performance*1 PDR−4J −20 to +100°C/5 to 98% rh (lowest attainable temperature in an ambient temperature of 0 to +30°C) Refer to diagram of temperature & humidity controllable range on page 26. Temp. & humidity fluctuation Temperature rate of change 1.5°C Heat up rate: 3.0°C/min. Heat up rate: 3.0°C/min. Pull down rate: 2.0°C/min. Pull down rate: 1.
PCR −20 to +100℃・30 to 90%rh CLEAN TEMPERATURE & HUMIDITY CHAMBER Model PCR−3J System Balanced Temperature and Humidity Control system (BTHC system) Performance*1 Temp. & humidity range −20 to +100°C/30 to 90% rh (lowest attainable temperature in an ambient temperature of 0 to +30°C) Refer to diagram of temperature & humidity controllable range on page 26. Temp. & humidity fluctuation ±0.5°C/±2.5% rh Temperature gradient 5.0°C Temperature variation in space 5.0°C Heat up rate: 1.5°C/min.
PU −40 to +100℃(+150℃/+180℃) LOW TEMPERATURE CHAMBER Model PU−1J PU−2J System PU−3J Temperature range −40 to +100°C (lowest attainable temperature in an ambient temperature of 0 to +30°C) Performance*1 Temperature fluctuation ±0.3°C Temperature gradient 3.0°C Temperature variation in space 1.5°C Heat up rate: 3.0°C/min. Pull down rate: 2.0°C/min. Temperature rate of change Temperature extremes achievement time Allowable heat load*2 Heat up time: from +20 to +100°C 30 min.
PG −70 to +100℃(+150℃/+180℃) ULTRA LOW TEMPERATURE CHAMBER Model PG−2J System PG−4J Balanced Temperature Control system (BTC system) Temperature range −70 to +100°C (lowest attainable temperature in an ambient temperature of 0 to +30°C) Performance*1 Temperature fluctuation ±0.3°C Temperature gradient 3.0°C Temperature variation in space Temperature rate of change 1.
TEMPERATURE & HUMIDITY CONTROL RANGE ● PR·PL·PSL 100 98 90 80 70 60 100 95 90 85 80 75 70 Relative humidity 50 % rh 40 60 Relative humidity 50 % rh 40 30 30 20 20 10 10 0 10 20 30 15 40 50 60 70 80 90 0 100 Temperature °C * Restrictions on continuous humidity operation at +40°C or lower because of frost on the cooler.
FITTINGS SAFETY DEVICES • Power cable connection port··········································· 2 • Control circuit overcurrent protection • Dew tray··········································································· 1 • Glass tube fuse for control circuit short-circuit protection • Drain hose (approx.
OPTIONS Can be equipped only at the time of unit purchase. Page OPTION PR PL PSL PHP PDR/PDL Can be added after unit is purchased. PCR PU PG Wide-view door *1 Door without viewing window Hand-in ports (For door with standard viewing window) *1 P.30 Hand-in ports (For door with Wide-view window) *2 Roller blind for wide-view window Inner glass door Power cable Power socket *3 Direct water coupling to tap water P.
OPTIONS Can be equipped only at the time of unit purchase. Page P.34 OPTION PR PL PSL PHP Temp. & humid. SP attainment output Program-synched DC power supply Paperless recorder Temperature (humidity) recorder Temperature (humidity) recorder wiring P.35 Recorder output terminal (temperature, humidity, and heater output) Recorder output terminal (dry [wet] bulb temperature) Thermocouple Wet bulb wick Power meter Web camera Folding table *1 Overcool protector Additional overheat protector P.
OPTIONS Wide-view door Effective view: Type 2 W470 x H720 mm Type 3 W570 x H820 mm Type 4 W970 x H970 mm * Standard performance may not be met under certain conditions. Inquire for details.
OPTIONS Power cable Water purifier (reverse osmosis) Water tank • 2.5 m •5m • 10 m * If this option is not specified, the chamber does not come with a power cable. Use to continuously supply pure water. • WS-1 Produced water capacity: 12 L/h (Water temperature: 25°C) Size: W400 × H400 × D280 mm For supplying water to the chamber’s fixed tank.
OPTIONS Shelf/shelf bracket Specimen basket Precision inner chamber < Shelf > • 18-8Cr-Ni Stainless steel For small specimens that cannot be placed directly on the shelf.
OPTIONS Additional cable port Cable port dew tray (for left side) DC inverter refrigeration Provided in addition/ replacement of the standard cable port (left side). Comes with a cap and a rubber plug. • ø25 mm (left-side or ceiling) Catches dew that comes out of the cable port. Can reduce power consu mpt ion when operating at low temperatures of 0°C or below as well as shorten temperature pulldown time.
OPTIONS Airflow adjuster Time up output Temp. & humid. SP attainment output Used when tests requi re low ai r f low velocity or a constant velocity. Setting value range: 4 levels This option enables turning the power to the specimen ON or OFF with contact signal output when the time is up by using the timer function on the temperature (humidity) controller. When the temperature (humidity) in the ch a mb e r r e a che s t he s e t va lue s , t he chamber sends out a contact signal.
OPTIONS Paperless recorder Temperature (humidity) recorder Recorder output terminal Records the temperature and humidity of each section such as the temperature inside the chamber. Data saving cycle: 5 sec. External recording media: CF memory card port (Includes a 256 MB CF card) USB memory port No.
OPTIONS Power meter Overcool protector Status indicator light Displays the integral power consumption for the chamber. If the temperature inside the chamber decreases excessively, the chamber stops operating to prevent the specimens from being damaged. Select light color, lighting, and blinking or buzzer sound.
OPTIONS Power key switch Anchoring fixtures Operation panel cover Used to manage/rest r ict the chamber usage. Used to fix the chamber to the floor. * Anchoring fixtures when installing the dew tray are also available. A cover for the operation panel. (Plastic) Chamber dew tray Prevents water leaks from the chamber onto the floor. Power indicator The operator can verify if the breaker is ON or OFF from the chamber front.
OPTIONS Test area low-silicone Air circulator removed for move-in Reduces the production of silicone gas (siloxane) in the test area. To prevent d a mage cau sed by heig ht restrictions, the air circulator for type 4 chambers is not mounted on the chamber during shipment. * The air circulator must be installed separately. Brake oil protection Changes resin par ts (water tan k front cover, door dew tray, chamber dew tray) to stainless steel.
http://www.espec.co.jp/english Head Office 3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan Tel : 81-6-6358-4741 Fax : 81-6-6358-5500 ESPEC NORTH AMERICA, INC. Tel : 1-616-896-6100 Fax : 1-616-896-6150 ESPEC EUROPE GmbH Tel : 49-89-1893-9630 Fax : 49-89-1893-96379 ESPEC ENVIRONMENTAL EQUIPMENT (SHANGHAI) CO., LTD.