9 Technical Specifications
46 |
46
201000401000 / Rev 04
9 Technical Specifications
9.1 1-Channel Transmitter
9.1.1 General Data
9.1.2 Electronic System
Description Data
Classification Device with internal power supply; type CF, defibrilla-
tion-proof applied part; class IIa
Number of ECG channels 1
Operating mode Continuous operation
Operating time on battery power Approx. 6 hours of continuous operation
Battery charging time 2.5 hours, charging current 500 mA, 1020 mAh recharge-
able battery capacity
Controls and indicators • LED indicators for operating status
• On/Off button
Biocompatibility To ISO10993
Respiration, leads-off sensing and
active noise suppression
No
Defibrillation protection Yes
Avg. recovery time < 5 s
Description Data
Battery type 1 x HR03/AAA NiMH battery, 1020 mAh
Amplitude response 0.05 to 125 Hz
AD converter 250 Hz, 12 bit (2.93 μV)
Input voltage range for ECG +/– 6 mV
Common mode rejection CMR > 80 dBm
Input impedance > 10 MΩ