Weighing Indicator User's Guide
    X K 3 1 9 0 - A 1
+P 
3. Calibration 
According to the Graph 2-3, connecting Indicator and loadcell, and the indicator enters 
weighing mode. 
1.  Connecting calibration jumper to the 15- pin socket on the indicator back panel. 
(There is a 15-pin plug in the packing carton with its 14-pin short connected with 
15-pin. ) 
2.  According to table 4-3, carry out calibration step by step. 
 Table 4 - 1 
step Operation  Display  Explanation 
1  press [ CALB]    after calibration jumper is inserted 
2 
press [1 ] [0 ] 
press [Enter ] 
[E ** ] 
[E 10 ] 
Enter division value selecting 
1/2/5/10/20/50/100/200 
Example: 10 
3 
press [ 0 ] 
press [ Enter ] 
[ dc * ] 
[ dc 0 ] 
Enter decimal point (0 -4) 
Example: without Decimal point 0 
4 
press [ 2 ] [ 3 ] 
press [ Enter ] 
[pon XY ] 
[ pon 23 ] 
Enter Zero range 
X: zero set range (1-5) it stands for 
 2% 4% 10% 20% 100% of F.S 
Y: power on Zero set range it 
 Stands for (1-5) 
 2% 4% 10% 20% 100% of F.S 
 example 23 
5 
press [ 3 ] [ 0 ] 
[ 0 ] [ 9 ] [ 0 ] 
[F ***** ] 
[ F 30090 ] 
Enter overload warning value when 
inputting F value,and re-pressing 
[ Enter ] key, the calibration will begin. 
directly pressing [ Enter ],will enter into 
the tenth step. If pressing [ weighting ], 
Indicator will return to weighing status. 
6  press [ Enter ]    Example: 30090 
7 
[ noLoAdn]  Confirming Zero position. At this time 
there is no load on the scale.Pressing 
[Input] when the stable light is on. 










