Weighing Indicator User's Guide
    X K 3 1 9 0 - A 1
+P 
3 Err 21 Calibrating data have been lost in RAM and EPROM, Operator must 
put 
 in the calibration plug, then reenter the original calibration data, turn 
 on the indicator again or re-calibrate it. 
4 Err 22 EPROM has been damaged. 
5 Err 23 RAM has been damaged. 
6.6 Other information indicating 
1. Err 24 In normal operation, you must pull off the calibration jumper.The 
indicator is under normal operation, and the calibration jumper shouldn’t be put in. If 
calibration is needed, the calibration jumper should be put in at set-up. You should pull 
out the calibration jump when turning off the indicator. If the indicator finds that the 
calibration jump was inserted when turning on, it will display Err 24  for 6 seconds 
to 
prompt the operator. 
 2 Err 25 Illegal software, or E2PROM was damaged 
. 
12 groups of date constitutes of a frame 
Date No. in a frame   explanation 
 1  02 (X ON) , start 
 2   + or - , 
 3   Weight date 
 4   Weight date 
 5   Weight date 
 6   Weight date 
 7   Weight date 
 8   Weight date 
 9   Decimal position, from right to left (0 – 4) 
 10   Checking XRL, higher 4 bits 
 11   Checking XRL, lower 4 bits 
 12   03 (X OFF) END 