User manual
  X K 3 1 9 0C 602   
and data format are same as Note 8 of E-2. 
Table E-2 Note 10: The meaning of NN parameter is shown in Table E-8. 
Table E-8 Definition of Parameters 
Parameter code  Definition 
TA~TH 
Remaining time of time relay 0 ~ time relay 7; integer 
of returned 3 bytes of C602 represents calculated value 
of  remaining  time.  Time  unit  100ms,  value  range 
0~255. 
JA~JH 
States  of group  0 to  group  7 intermediate relays, 8 
intermediate  relays  for  every group.  Group 0  is  time 
relay. 8 digits of binary data represented by returned 3 
digits of integer of C602 indicate the state of 8 relays. 
PA~PZ 
26  memories,  including  integers  ranged 
-8388608~8388607; long integer of returned 8 bytes of 
C602 
Note 11 of Table E-2: this command is only effective in remote control state. Parameter definition is shown 
in Table E-7. Time relay state can not be written. 
Table E-2 Note 12: returned character string is the content of corresponding relay.