User manual
  X K 3 1 9 0C 602   
Send  by 
PC 
Send IO state 
02H  Addr  AC  NN 
****** XH XL 03H 
0
2H    AACO0013D 
03H 
Note 9 AC
Send  by 
instrument 
Send IO state 
02H  Addr  AC  NN 
****** XH XL 03H 
02H    AACO0013D 
03H 
Send  by 
PC 
Read internal 
memory state 
02H Addr  AD NN XH 
XL 03H 
02H AAD J13F 03H 
Note 10 
AD
Send  by 
instrument 
Send internal 
memory state 
02H  Addr  AD  NN 
****** XH XL 03H 
02H    AADJ1******** 
03H 
Send  by 
PC 
Write internal 
memory 
02H  Addr  AE  NN 
****** XH XL 03H 
02H AAEJ1013F 03H
Note 11 
AE
Send  by 
instrument 
Send internal 
memory state 
02H  Addr  AE  NN 
****** XH XL 03H 
02H AAEJ1013F 03H
Send  by 
PC 
Read auxiliary 
display 
02H  Addr  AF XH  XL 
03H 
02H AAF46 03H 
AF
Send  by 
instrument 
Send auxiliary 
display 
02H  Addr  AF  ****** 
XH XL 03H 
02H    AAF******** 
03H 
Note 12 
Send  by 
PC 
Read the 
content of main 
display 
02H  Addr  AG XH  XL 
03H 
02H AAG46 03H 
AG
Send  by 
instrument 
Send the 
content of main 
display 
02H  Addr  AG  ****** 
XH XL 03H 
02H    AAG******** 
03H 
Note 12 
Note 1 of Table E-2: Instrument address is supposed to be 1, represented by letter “A”. If instrument address 
is  not  “1”,  checksum  XH  and  XL  should  be  changed  accordingly.  If  response  character  string  includes  “en” 
character, it means the format of this command is wrong or execution condition is not satisfied. 
Note 2 of Table E-2: The codes of calibration parameter are: e verification scale interval: Dp decimal digit; F 
maximum capacity; B1
 CAL coefficient; 0P zero point AD code; ON current zero point (read only); NL non-linear 
modification  value;  AD  AD  transfer  speed;  FL  filtration  strength;  Fm  filtration  calculation  method;  St  stable 
judgment range; 0T zero point tracking range; 0S zero setting range; 0I initial zero setting scope; EI Switch of 
zero-setting upon startup; Ut measuring unit. During calibration parameter modification, calibration switch must 
be turned on. See Table-3 for returned data and explanation. 
Table E-3 Explanation of Returned Calibration Data of Instrument 
Returned data of 
instrument 
Explanation 
AQe 01 54  Scale intervaleffective digit 1 54 is check 
sum, same below 
AQDp 3 17  Decimal digit 3 
AQF 020.000 7A  Maximum capacity 20kg 
AQBl 0067106 28  Standard rate 
AQ0P 0262122 67  AD code of calibration zero 
AQ0N 000.012 **  Current zero (  deviation relevant to  calibration 