Display method of program step
 Set Parameter 17B ofSET 0to 4. Under running status, 
the auxiliary display of  the instrument  will show the current program step number in forms of 
‘Step**’.
Table 8-4 Description of Program Step in External Control Mode 
Program 
Step 
Description 
1 
If the external signal I3 is valid, go to program step 2 and the timer 
will be T0. 
2 
Timing  T0  is  over.  Start  to  calculate  the  average  weight,  go  to 
program step 2 and the timer will be T1. 
3 
Record the weight record. Timing T1 is over; calculate the average 
weight during the time interval T1, go to program step 4 and the 
timer will be T2. 
4 
Timing T2 is over. Save the weight data and sort it by the weight. 
Output the batching signal by the channel and timer will be T3. 
5 
Timing T3 is over. Close the batching signal. If the weight is back 
to zero zone, go to program step 1. 