off quick feed. The timer will be T1 and go to program step 3. 
3  The timing is over, go to program step 4 and the timer will be T0. 
4  Start the slow feed; timing is over and weight inspection is started. 
If the loaded value reaches to batching value minus slow feed lead, 
turn off slow feed. Timer will be T2 and go to program step 5. 
5  The timing is  over.  If the lead correction is  allowed, modify  the 
slow feed lead and go to program step 6. 
6  If the weight is qualified, the qualified signal will be output and it 
will go to program step 9, the timer will be T5. If it is out of the 
tolerance, output the out-of-tolerance signal. If the weight is less 
than  the  batching value  minus  tolerance  and  the  gradual feed  is 
allowed,  go  to  program  step  7  and  the  timer  will  be  T3;  If  the 
gradual feed  is  not  allowed  and  the  weight  is  more  than  the 
batching value plus tolerance, out-of-tolerance treatment will stop 
at this step and wait for qualified out-of-tolerance treatment. If the 
out-of-tolerance treatment is not done, then go to program step 9, 
the timer will be T5. 
7  Start the slow feed, timing is over, the slow feed will be closed and 
timer will be T4. Go to program step 8. 
8  Treatment is same as program step 6 
9  Timing  is  over,  close  the  qualified  or  out-of-tolerance  signal.  If 
material feeding action is started manually, then stop the action; If 
it is under the status of automatic cycle, then go to program step 
10. 
10  Perform  the  automatic  print  and  automatic  storage  operation 
according to the setting and go to the program step 11. 
11  Exit the  status of  negative scale, go  to  program  step  12  and the 
timer will be T7. 
12  The  timing  is  over.  If  the  cycle  time  is  not  reached, go  back  to 