User manual
  X K 3 1 9 0C 602   
8 
[******]
[0Point] 
Current Zero 
(
Relative  to  calibration 
zero) 
Can
not  be  modified. 
This parameter
  reflects 
the  z
ero  stability  of 
weighing load cell. 
9 
[  A]
[Print?] 
A—
Whether to print each 
parameter 
0No1Yes   
Note 1:  Nonlinear correction uses the parabola calculation method.  The correction  value is  1/2  max. weighing 
position, that is, the correction amount at the top of the correction curve. Nonlinear correction value equals to the 
negative value of the nonlinear tolerance of 1/2 max. weighing position. The unit of the nonlinear correction value 
is 1% of the max. weighing value. For example, for  the scale whose  Max=10kg, if 5kg is  added to  it and the 
weight shows 5006g, then the tolerance is +6g, that is , +0.06%Max and the nonlinear correction value is -0.06. 