[t0 ]
Judgment delay (0 ~ 25.5)
seconds: 
U
nder external control mode, 
when  it  is  activated, 
data 
calculation  is  started 
after 
t0; 
Under self-
control mode, data 
calculation  is  started 
after  t0 
when weight is  out of
  zero 
zone. 
Please  refer  to  p
arameter  6 
for  the  definition  of  โ
zero 
zoneโ. 
8 
[ ***]
[t1 ] 
T
ime  for  calculation  of 
average weight 
(0 ~ 25.5) seconds: 
After  t0,  the  indicator
  will 
sum  the weight  values  and 
figure  out  the  average 
during time t1. 
The obtained 
value  serves  as
catchweighing basis.  
A
ssume  that  the  obtained 
value is X. 
9 
[ ***]
[t2 ]
Calculation delay (0 ~ 
25.5) 
seconds: 
After t1
 calculation is over, 
the indicator
 remains inactive 
during  t2  time,  waiting
  with 
delay. 
T
his setting can meet  more 
control site requirements. 
10 
[ ***]
[t3 ]
Time for sending signal   
(0 ~ 25.5) seconds: 
After  t2,  the  indicator
  sends 
catchweighing  signal  with
length of t3 
P
lease  refer  to  the 
explanation  of  parameter  2
~ parameter 5. 
11 
[  A]
[Print?] 
Aโ
whether  to  print 
parameter setting 
(0: No: 1: Yes) 
II. Calibration    
Message: The calibration switch should  be opened  when the calibration is  started, 
otherwise the calibration status cannot be entered. Before exit by pressing the button [enter], 
the  calibration  switch  should  be  closed  after  completion  of  calibration;  otherwise  the 
auxiliary display will show the message [CALEnd]; It can be quitted by pressing [exit] and 