User manual
  X K 3 1 9 0C 602   
Chapter 4 Parameters Setting and Calibration 
I. Parameters setting 
The indicator has 5 groups of parameters setting condition, which are as follows: 
SET 0: inquiry parameters;   
SET 1:general parameters; 
SET 2: control parameters; 
SET 3: calibration; 
SET 4: calibration parameters. 
  Press setting button to enter   , press 5or 6to select the parameter group, 
then pressinput button to enter the setting of relevant parameter. 
  Some setting can go into effect after re-startup and power on. 
Attention:  when  it  is  necessary  to  modify  the  calibrated  parameters,  the  calibration 
switch  shall  be  turned  on.  Or  else,  the  modification  cannot  be  carried  out  and  warning 
“Error 7” will be prompted. 
The contents of parameter table 
SET 0
 and
SET 2
 depend on the parameter 
13 A in parameter table 
SET 1
. 
Initialization of parameters 
 During  the  calibration  process  of  (SET  3),  input  “123456”  (ignore  the  decimal)  while 
inputting the loaded weight, pressinput button, the indicator will carry out initialization of 
parameters, all calibration parameters and working parameters will go back to the factory 
setting. Under normal situation, special attention should be paid when using this function.   
Explanation of button in parameters setting 
7or8 change the currently flickering digit; 
5or 6modify the value (parameter) of the currently flickering digit; 
esc  do not save the current modified parameter, go back to the weighing status. 
input  if a parameter is modified, pressing this button will save the parameter and enter the 
[ 0] 
[SEt ]