User manual

X K 3 1 9 0C 602
Chapter 4 Parameters Setting and Calibration
I. Parameters setting
The indicator has 5 groups of parameters setting condition, which are as follows:
SET 0: inquiry parameters;
SET 1:general parameters;
SET 2: control parameters;
SET 3: calibration;
SET 4: calibration parameters.
Press setting button to enter , press 5or 6to select the parameter group,
then pressinput button to enter the setting of relevant parameter.
Some setting can go into effect after re-startup and power on.
Attention: when it is necessary to modify the calibrated parameters, the calibration
switch shall be turned on. Or else, the modification cannot be carried out and warning
“Error 7” will be prompted.
The contents of parameter table
SET 0
and
SET 2
depend on the parameter
13 A in parameter table
SET 1
.
Initialization of parameters
During the calibration process of (SET 3), input “123456” (ignore the decimal) while
inputting the loaded weight, pressinput button, the indicator will carry out initialization of
parameters, all calibration parameters and working parameters will go back to the factory
setting. Under normal situation, special attention should be paid when using this function.
Explanation of button in parameters setting
7or8 change the currently flickering digit;
5or 6modify the value (parameter) of the currently flickering digit;
esc do not save the current modified parameter, go back to the weighing status.
input if a parameter is modified, pressing this button will save the parameter and enter the
[ 0]
[SEt ]