  X K 3 1 9 0C 602   
Figure 3-9 Setting of Analog Output 
Calibration method of Analog output 
  The  zero  point  value  and  full  range  value  of  analog  quantity  are  proportional  to  the 
corresponding DA code (see table 4-3, explanation forSET 1 parameter 11 and 12). Calculation 
and calibration of parameter 11 and 12 can be done by using the tolerance of Analog output. 
Attention: in voltage output mode, short circuit of Analog output terminal is strictly 
prohibited, so does the connection of load of 4-20mA current output. Or else, damage will 
be caused to the Analog output circuit. 
X. Detection of input and output hardware 
When the  main display  is  in  the  display  internal  code  mode (select 1  for  parameter A of   
in SET 0), the detection can be done to check the condition of input and output 
terminals. When I0 input terminal signal is effective, the corresponding O0 will have 
control  signal  output.  For  the  same  reason,  the  I1  corresponds  to  O1……I7 
corresponds to O7. 
Attention:  when  external  equipment  is  connected,  try  to  avoid  using  this  method  to 
detect input and output circuit so as to prevent any accident. 
Under the displaying status of internal code, the indicating lamp for net weight reflects the 
working status of ARM single-chip computer oscillator in main panel. When the indicating lamp 
is on,  it  means that  the  external quartz resonator is  operating normally  while the  off status of 
indicating lamp means that the internal RC oscillator within single-chip computer is working.   
[ AB] 
[DISP ] 