User manual
  X K 3 1 9 0C 602   
Figure 3-8 Switch input and Output Interfaces 
Attention: among the input and output terminals, it is prohibited to short 0V terminal 
and +12V terminal. Or else, severe damage will be done to the indicator. 
IX. Analog output 
C602 indicator can choose from three Analog output modes: 0 ~ 5V, 0 ~ 10V voltage signal 
output  and  4  ~  20mA  current  signal  output.  The  current  loop  is  provided  by  internal  power 
supply and the current signal can be adjusted to 0 ~ 20mA. The selection of output mode can be 
done in the jumper JP1 ~ JP3 in power panel, see figure 3-9 for details. In the figure, the jumper 
position is in conformity with these in power panel and X stands for the position of short circuit 
ring. The factory setting of current loop is 4-20mA. The output value can be gross weight or net 
weight by setting SET 1 parameter 1F. The analog output switch is controlled by SET 1
parameter 1E. 
Jumper selection 
Analog output 
JP3  JP2  JP1 
    X 4—20mA 
(factory setting) 
X  X   
X     
0—5V 
  X  X 
X  X   
0—10V 
    X 
Definition of 
catch- 
weigher 
terminal 
S 
t 
a
n
d
b
y 
S
t
a
n
d
b
y 
C
h
a
n
n
el 
5 
C
h
a
n
n
e
l 
4 
C
h
a
n
n
el 
3 
C
h
a
n
n
e
l 
2 
C
h
a
n
n
el 
1 
O
p
e
r
a
ti
o
n 
S
ta
n
d
b
y 
S
t
a
n
d
b
y 
S
ta
n
d
b
y 
S
t
a
n
d
b
y 
E
x
te
r
n
al 
c
o
n
tr
o
l 
S
t
o
p 
Oper
ation 
Standby 
C
orresponding 
indicating 
lamp on panel
  O7
O6
O5
O4
O3
O2
O1
O0
I7  I6  I5  I4  I3  I2  I1  I0   










