III. Connection between load cell and indicator 
1. 7-pin electrical male plug are adopted for connection of load cell. Figure 3-5 shows the 
meanings of each pin. 
2. When junction box is used to connect several load cells or extend the cable of load cell, 
six-wire  connecting  method  must  be  adopted.  When  1  load  cell  is  used  and  the  cable  is  not 
extended, four-wire system can be adopted. Under this condition, it is necessary to short connect 
the EX+ and SEN+ with EX- and SEN- respectively. 
 3. The load cell and indicator should be reliably connected and the shielded line of load 
cell should be reliably connected with GND terminal. The connecting line can not be plugged 
in and out when the indicator is powered on in order to prevent any damage to the indicator 
or load cell by static electricity 
 4. The load cell and indicator are both static sensitive equipments, so anti-static measures 
must be taken during the use. It is strictly forbidden to carry out welding operation or other 
operations  with  high  current  on  the  weighing  platform.  In  the  stormy  season,  lightening 
prevention measures must be taken reliably to prevent any damage to load cell and indicator 
caused by lightening stroke, and to guarantee the personal security of operators and safe 
running of weighing devices and relative equipments. 
In the figure: EX+, EX-: excitation power supply; SEN+, SEN-: excitation feedback; SIG+, SIG-: load cell output 
signal; GND: shield 
 Figure 3-5 Connection Diagram of Load cell 