Figure 5: Gantry starshot film image and selection of vertical fiducial mark
Figure 6: Gantry starshot with software fiducial marks in place
6. Click on “Tool – starshot physics QA” to activate the analysis. The
fiducial marks should automatically align with the axes in the display, but if
they do not as shown in Figure 7 click on the . icon and select “Fit
fiducials horizontally, vertically and rotationally”. The aligned image is
shown in Figure 8.