Here’s where the One Scan Protocol can be
applied. Display the image of the application
and reference films in the Image Panel and
draw an area of interest on the unexposed
reference strip. Then right click the region to
designate it a calibration region and right click
again on the frame to type in and enter the
calibration value of zero. Notice that the frame
changes color from yellow to blue to indicate
it’s a calibration region. Repeat the process
for the exposed strip and enter the value of
313 cGy. Now right click on the line “Dose
Map (Single Scan)” and select “Dose Mapping
Method” then “Recalibration using reference
regions” and then “Dose Linear Scaling”. To
complete the recalibration, go back to “Dose Map (Single Scan)” and select “Rebuild
Dose Map”. The thumbnail images of the Dose and Consistency Maps are refreshed.