Note: For absolute dosimetry, do NOT adjust the maximum dose in the plan image.
Maximum dose could be adjusted when doing relative dosimetry.
Small position adjustments of x = - 0.3 mm; y = -0.6mm and rotation by -0.2° move the
passing rate from 95.11% to 96.4%. Clicking the icon assigns the analysis map
(gamma map in the example) as a Treatment Object within the Dose to plan
comparison Object. Note: If the analysis function values do not update after changes
click the icon to force an update.
The icons under the
histogram in the
Analysis Window control
the range of values/step
sizes along the x axis of
the chart (the gamma
value axis in the
example). Click to
change the way the
histogram is displayed.
The red color channel
icon, indicates data for the red color channel is displayed. Click to change the color
channel.