User manual

45
User defined dose range: Scales the dose range of the dose map to the
user’s values – this is helpful when the film image contains dark areas not
related to dose e.g. pen marks, scanner masks
Enforce same dose range for RGB: When checked, the dose range is the
same for all color channels
Uniformity Map: Click this line to calculate and display the uniformity map. In
triple channel dosimetry, the film image is split into dose-dependent and dose-
independent parts. The uniformity map is the dose-independent part and
contains the film and scanner artifacts removed from the dose map.
Consistency Map: Click this link to access options for displaying and scaling the
consistency map. In triple channel dosimetry, the consistency map is the
remaining error after the dose-independent information is removed. Lower
consistency means a better calibration.