DOSE MAP (Video: http://www.youtube.com/watch?v=pfcGM3yIHlU)
To measure or evaluate a dosimetry film, it is necessary to first scan the film and apply
an appropriate calibration function to convert the resulting image into dose space. Once
converted to a dose image, FilmQA
™
Pro offers an array of quantitative analysis tools to
evaluate the dose maps.
To work with an exposed
application film begin by
opening a Calibration Object
and developing a calibration
function. Alternatively open a
saved case containing a
calibration object/calibration
function or load a Calibration
Object or Fixed Calibration from
file. For either of the last two,
right click on “Add new case
object” in the Case Data
Selector window and from the
drop-down menu, select “Open
case object from file” then navigate to the correct folder and load the calibration file.
Note: The calibration function is scanner dependent as well as being dependent on the
film type and manufacturing lot. The film image being used for dose map calculation
must be acquired from the same scanner as the calibration data and be the same
acquisition type, e.g. reflection or transmission mode and positive or negative image. In
addition the application film must come from the same manufacturing lot as the
calibration film.