User manual
29
There are two ways to introduce calibration data or images into FilmQA
™
Pro. Film
Calibration (Ordinary) is used when all the calibration films are contained in a single
scan as for the One Scan Protocol. Film Calibration (Mosaic) is used when the
calibration films are contained in multiple images.
In the “Case Data Selector” window, select “Add New Case Object” under the “Case
Object Management” heading. From the drop-down menu choose either “Film
Calibration (Ordinary)” or “Film Calibration (Mosaic)”. Both options are described below:
Film Calibration (Ordinary)
1. Click the “+” sign next to “Film Calibration
ordinary)” to expand the selection, right click on
the first heading on the branch “Data – N/A
(Empty)” and select from the menu:
“Open image ‘Calibration film’ from file”:
navigate to a folder and select a “tiff”
image file to open;
“Open image ‘’Calibration film’ from
DICOM file”: navigate to a folder and
select a DICOM file to open;
“Recent image files”: provides a shortcut
to recently used image files;
“Scan image ‘Calibration film’”: Scan and obtain an image from a Twain
compliant scanner (Video: http://www.youtube.com/watch?v=SfyaQl6rHdM)
2. Click on “Open image ‘Calibration film’ from
file”. Navigate to the
FilmQAPro/Images/Example EBT3 Rapid Arc
and open the calibration film image. A
thumbnail image appears in the empty data
slot and the full image is displayed in the
center window.