User manual
15
“Case Data” opens a drop down list to select case specific information related to
the patient. Double-click the icon to open the pop-up window and select the
preferred case. “Patient Data” will change to match the patient name of the
treatment case selected.
Case Object Management allows objects to be added to the Case Tree to create a film
dosimetry case. Under “Add New Case Object”:
“Film Calibration (Ordinary)” – Scan or
open a file containing a single image to
generate a calibration curve
“Film Calibration (Mosaic)” – Scan or
open multiple image files to generate a
calibration curve
“Film Calibration Dose Maps” –
Calculates a dose map, consistency
map and uniformity map from the image
or images used to generate the
calibration curve
“Images Only (Multiple Scans)” – Scan
or open multiple image files. The images can be averaged or fused together and
the resultant image can be saved or copied to be used as required.
“Dose Map (Single Scan)” – Scan or open an image file for conversion to a dose
map from a Film Calibration Object. A calibration curve is required in the Case
Tree before a dose map can be generated. Note: By default the first Film
Calibration Object in the case is used. In the Dose Map branch of the Case Tree
select a different dependency if the case contains multiple Calibration Objects
and one of the subsequent ones is to be used.
“Dose to Plan Comparison” – Import a treatment plan from a file, TPS or DICOM
and do a gamma analysis against a treatment film. A Dose map Object is
required in the Case Tree before a dose-to-plan comparison can be done. Note: