User manual
119
Figure 3A Figure 3B Figure 3C
7. Place the calibration films and an unexposed film from the same lot on the
scanner as shown in Figure 4. The time between film exposure and scanning is
related to the time window within which the calibration strip and application film
were exposed. Your efficiency increases when you minimize the time window.
For exposures t min. apart, film scanning can be done 4t min. later, or any time
thereafter.
Figure 4
Incorrect! Color correction
must be turned off.
Open Configuration window
and check “No color
correction”
Correct! Color
correction is not active
Scan Direction
Directiontionecti
on
Calibration Region
Keep first 2 cm clear of
film
Position exposed areas on
the film in the center (left-to-
right) of the scan window
with the 8” edge
perpendicular to the scan
direction
Start of scan