Specifications

1-24
8719/20/22ES Specifications and Characteristics
Instrument Specifications
Instrument Specifications
Uncorrected Port Performance
Table1-213.5-mm Device Connector Type
8719ES/20ES (3.5-mm, 50)
Description
Characteristic
50 to 500 MHz 0.5 to 2 GHz 2 to 8 GHz 8 to 20 GHz
Directivity
a
a.Does not include the effect of the cable set on the test ports.
24 27 21 16
Source Match
a
Standard (dB) 12 12 10 8
Option 400 (dB) 20 20 11 10
Option 007 (dB) 16 20 14 11
Option 085 (dB) 16 18 14 8
Load Match
a
Standard (dB) 22 20 12 10
Option 400 (dB) 20 17 12 10
Option 007 (dB) 26 24 15 12
Option 085 (dB) 26 24 15 10
Reflection Tracking
b
(dB)
b.Excludes rolloff below 500 MHz, which is typically18 dB at 100 MHz, and25 dB at
50MHz.
±3 ±3 ±3 ±3
Transmission
Tracking
b
(dB)
±3 ±3 ±3 ±3
Tracking Stability
(Ratio Measurement)
Magnitude (dB) 0.02/°C, typ. 0.03/°C, typ. 0.03/°C, typ. 0.04/°C, typ.
Phase (dB) 0.1/°C, typ. 0.1/°C, typ. 0.3/°C, typ. 0.50/°C, typ.
Crosstalk
c
(dB)
c.Measurement conditions: Normalized to a thru, measured with two shorts, 10 Hz IF
bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of the
maximum power out or the maximum receiver power.
75
d
d.Limited by noise floor.
95 91 86