Specifications
6-24
Error Messages
Error Messages in Alphabetical Order
SYST CTRL OR PASS CTRL IN LOCAL MENU
Error Number
36
The analyzer is in talker/listener mode. In this mode, the analyzer cannot
control a peripheral device on the bus. Use the local menu to change to system
controller or pass control mode.
TEST ABORTED
Error Number
113
You have prematurely stopped a service test.
TEST PORT OVERLOAD, REDUCE POWER
Error Number
57
You have exceeded approximately +14 dBm at one of the test ports (or 0 dBm at
the A or B sampler, Option 012 only). When this occurs, reduce the power to a
lower level.
THIS LIST FREQ INVALID
Error Number
133
You have set frequencies in the list that are outside of the allowable frequency
range of the analyzer. Reduce the frequency range of the list.
TOO MANY NESTED SEQUENCES. SEQ ABORTED
Error Number
164
You can only nest sequences to a maximum level of six. The sequence will abort
if you nest more than six.
TOO MANY SEGMENTS OR POINTS
Error Number
50
You can have a maximum of 30 segments or 1601 points in frequency list mode.
In power meter calibrations, you can have a maximum of 12 segments for power
sensor cal factors and power loss functions.
TRANSFORM, GATE NOT ALLOWED
Error Number
16
You can perform a time domain transformation only in linear and CW sweep
types.
TROUBLE! CHECK SETUP AND START OVER
Service Error
Number 115
Your equipment setup for the adjustment procedure in progress is not correct.
Check the setup diagram and instructions in the service guide. Start the
procedure again.