Specifications
6-15
Error Messages
Error Messages in Alphabetical Order
MEMORY FOR CURRENT SEQUENCE IS FULL
Error Number
132
All the memory in the sequence you are modifying is filled with instrument
commands.
MORE SLIDES NEEDED
Error Number
71
When you use a sliding load (in a user-defined calibration kit), you must set at
least three slide positions to complete the calibration.
NO CALIBRATION CURRENTLY IN PROGRESS
Error Number
69
The softkey is not valid unless a calibration is
already in progress. Start a new calibration.
NO DISK MEDIUM IN DRIVE
Error Number
41
You have no disk in the current disk unit. Insert a disk, or check the disk unit
number stored in the analyzer.
NO FAIL FOUND
Service Error
Number 114
The self-diagnose function of the instrumentoperates on an internaltest failure.
At this time, no failure has been detected.
NO FILE(S) FOUND ON DISK
Error Number
45
No files of the type created by an analyzer store operation were found on the
disk or the disk drive is empty. If you requested a specific file title, that file was
not found on the disk.
NO IF FOUND: CHECK R INPUT LEVEL
Error Number
5
The first IF signal was not detected during pretune. Check the front panel R
channel jumper. If there is no visible problem with the jumper, refer to the
service guide for troubleshooting.
NO LIMIT LINES DISPLAYED
Error Number
144
You can turn limit lines on but they cannot be displayed on polar or Smith chart
display formats.
NO MARKER DELTA - SPAN NOT SET
Error Number
15
You must turn the delta marker mode on, with at least two markers displayed,
in order to use the softkey function.
RESUME CAL SEQUENCE
MARKER
→
SPAN