Specifications
6-12
Error Messages
Error Messages in Alphabetical Order
FREQ OFFSET ONLY VALID IN NETWORK ANALYZER MODE (ES Option 089 Only)
Error Number
140
You can only make frequency offset measurements in the network analyzer
mode.
FREQS CANNOT BE CHANGED, TOO MANY POINTS
Error Number
204
The number of points selected for setting the low pass transform frequencies is
too high. Reduce the number of points so that the low pass criteria is met.
FUNCTION NOT AVAILABLE
Error Number
202
The function you requested over GPIB is not available on the current
instrument.
FUNCTION NOT VALID
Error Number
14
The function you requested is incompatible with the current instrument state.
FUNCTION NOT VALID DURING MOD SEQUENCE
Error Number
131
You cannot perform sequencing operations while a sequence is being modified.
FUNCTION NOT VALID FOR INTERNAL MEMORY
Error Number
201
The function you selected only works with disk files.
FUNCTION ONLY VALID DURING MOD SEQUENCE
Error Number
163
You can only use the capability when you are building a
sequence. Attempting to use this softkey at any other time returns an error
message and no action is taken.
8720 SOURCE PARAMETERS CHANGED
Error Number
61
Some of the stimulus parameters of the instrument state have been changed,
because you have turned correction on. A calibration set for the current
measurement parameter was found and activated. The instrument state was
updated to match the stimulus parameters of the calibration state. This
message also appears when you have turned on harmonic mode or frequency
offset, and the present frequency range cannot be used with one of these modes.
GOSUB SEQUENCE