Specifications

1-10
8719/20/22ES Specifications and Characteristics
Corrected System Performance (8719/20ES)
Table 1-8 7-mm Device Connector Type
Network Analyzer: 8719ES/20ES, Standard
Calibration Kit: 85050B (7-mm, 50 )
Cables: 85132F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8
Description Specification
50 to 500 MHz 0.5 to 2 GHz 2 to 8 GHz 8 to 18 GHz
Directivity (dB) 52 52 52 52
Source Match (dB) 48 48 44 41
Load Match (dB) 51 51 51 51
Refl. Tracking
Magnitude (dB) ±(0.006 + .02/°C) ±(0.006 + .03/°C) ±(0.017 + .03/°C) ±(0.047 + .04/°C)
Phase (deg) ±(0.040 + 0.1/°C) ±(0.040 + 0.1/°C) ±(0.112 + 0.3/°C) ±(0.310 + 0.5/°C)
Trans. Tracking
Magnitude (dB) ±(0.01 + .02/°C) ±(0.011 + .03/°C) ±(0.022 + .03/°C) ±(0.034 + .04/°C)
Phase (deg) ±(0.066 + 0.1/°C) ±(0.073 + 0.1/°C) ±(0.145 + 0.3/°C) ±(0.224 + 0.5/°C)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)