Specifications
5-49
Hardkey/Softkey Reference
Analyzer Functions
defines the measurement as S
22
, the complex reflection
coefficient (magnitude and phase) of the test device
output.
measures the reflection and thru paths of the current
calibration standard.
leads to the reflection calibration menu.
leads to the reflection calibration menu (ET only). In the
ECal menu, this softkey performs the reflection 1-port
calibration.
is located in the ECal service menu. This allows you to
select different reflection standards in the ECal module.
The number of reflection standards varies with ECal
module type.
completes the adapter removal procedure, removing the
effects of the adapter being used.
allows you to change the name of a file that has already
been saved.
resets the color being modified to the default color.
• When in the specify class more menu, is
used to enter the standard numbers for a response
calibration. This calibration corrects for frequency
response in either reflection or transmission
measurements, depending on the parameter being
measured when a calibration is performed. (For default
kits, the standard is either the open or short for
reflection measurements, or the thru for transmission
measurements.)
• When in the response cal menu, leads to
the frequency response calibration. This is the simplest
and fastest accuracy enhancement procedure, but
should be used when extreme accuracy is not required.
It effectively removes the frequency response errors of
the test setup for reflection or transmission
measurements.
• When in the specify class more menu,
is used to enter the standard
numbers for a response and isolation calibration. This
calibration corrects for frequency response and
directivity in reflection measurements, or frequency
response and isolation in transmission measurements.
REFL: REV S22 (B/R)
REFLECT AND LINE
REFLECTION
REFLECTION 1-PORT
REFL STD [0]
REMOVE ADAPTER
RENAME FILE
RESET COLOR
RESPONSE
RESPONSE
RESPONSE
RESPONSE ISOL'N
RESPONSE ISOL'N